ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2012-10932
- Event Type
- Malfunction
- Date Received
- September 20, 2012
- Date of Event
- August 24, 2012
- Report Date
- August 24, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
FOLLOW UP #1 SUBMITTED: (B)(4) 2012. DEVICE EVALUATION: A RETAINED CARTRIDGE SAMPLE FROM LOT # B201746 WAS EVALUATED. A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A LEAK TEST, FILL TEST, AND FORCE TEST WAS PERFORMED WITH NO FAILURES OBSERVED. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.
THE CARTRIDGES HAVE NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2012 THE PATIENT CONTACTED ANIMAS ALLEGING THAT THE CURRENT BOX OF CARTRIDGES SHE IS USING HAVE ALL LEAKED FROM THE PLUNGER END OF THE CARTRIDGE SINCE THE BOX WAS OPENED FOR US. THE PATIENT STATED THAT 5 CARTRIDGES FROM THE SAME BOX HAVE BEEN USED, AND ALL HAVE LEAKED FROM THE PLUNGER END. THE PATIENT DENIED ANY DAMAGE TO THE CARTRIDGES. THE PATIENT NOTED THAT THE CURRENT CARTRIDGE SHE HAD IN THE PUMP WAS NOT LEAKING. THERE WAS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED CARTRIDGE ISSUE REMAINED UNRESOLVED AND BECAUSE A LEAKING CARTRIDGE CAN LEAD TO UNDER-DELIVERY OF INSULIN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN CARTRIDGE | INSULIN PUMP CARTRIDGE | LZG | ANIMAS CORPORATION | ANM IR1200/1250/2020/OTP CART | B201746 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |