ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2012-00646
- Event Type
- Malfunction
- Date Received
- September 19, 2012
- Date of Event
- August 7, 2012
- Report Date
- August 21, 2012
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- K983112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE FOR EVALUATION. THE DRYLINE WAS VISUALLY INSPECTED FOR THE REPORTED DAMAGE. RESULTS: THE VISUAL INSPECTION IDENTIFIED A HOLE APPROXIMATELY 10CM FROM THE CONNECTOR ON THE RETURNED DRYLINE TUBE. A LOT CHECK REVEALED SEVEN OTHER SIMILAR COMPLAINTS FOR THE LOT NUMBER 120127. CONCLUSION: IT WAS IDENTIFIED THAT THE DAMAGE TO THE RT340 DRYLINE COULD HAVE BEEN CAUSED DURING THE MANUFACTURING PROCESS. DURING THE PRODUCTION OF THE DRYLINE TUBES, A PRESSURE TEST IS PERFORMED EVERY 10 TO 15 MINUTES AND THOSE THAT FAIL ARE SET ASIDE FOR FURTHER INSPECTION. THE USER INSTRUCTIONS SUPPLIED WITH THE RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT STATE THE FOLLOWING: -PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT; -SET APPROPRIATE VENTILATOR ALARMS; (B)(4).
A HOSPITAL IN (B)(6) REPORTED THAT AN RT340 ADULT BREATHING CIRCUIT FAILED THE LEAK TEST ON THE RESPIRATOR. THEY FURTHER STATED THAT A PIN HOLE WAS FOUND ON THE DRYLINE . THIS WAS FOUND BEFORE PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT340 | 120127 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DRAGER EVITA 4 RESPIRATOR |