ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2012-00644
- Event Type
- Malfunction
- Date Received
- September 18, 2012
- Date of Event
- July 27, 2012
- Report Date
- August 22, 2012
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- K983112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT RT340 BREATHING CIRCUITS WERE NOT RETURNED TO FISHER & PAYKEL HEALTHCARE (FPH) (B)(4) FOR EVALUATION. OUR INVESTIGATION IS THEREFORE BASED ON THE DESCRIPTION OF THE PROBLEM AND OUR KNOWLEDGE OF THE PRODUCT. THE THREE COMPLAINT BREATHING CIRCUITS WERE FROM LOTS 120404, 120425 AND 120523 (MANUFACTURED ON 4 APRIL 2012, 25 APRIL 2012 AND 23 MAY 2012 RESPECTIVELY). RESULTS: BASED ON OTHER COMPLAINTS OF THIS NATURE THAT WE HAVE RECEIVED, WHEN THE HEATER WIRE ADAPTOR CANNOT BE FULLY CONNECTED TO THE HEATER WIRE SOCKET IN THE INSPIRATORY TUBE OF THE RT340, WE USUALLY FIND THAT ONE OR BOTH OF THE INSPIRATORY HEATER WIRE PINS ARE BENT OR SPLIT. THIS PREVENTS THE ADAPTOR FROM CONNECTING TO THE HEATER WIRE SOCKET. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR THE THREE LOT NUMBERS PROVIDED. CONCLUSION: ALL BREATHING CIRCUITS ARE TESTED FOR CONNECTIVITY AND ELECTRICAL CONTINUITY DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. IT IS POSSIBLE FOR THE USER TO BEND OR SPLIT THE HEATER WIRE PINS DURING USE IF THE HEATER WIRE ADAPTOR IS INSERTED INTO THE HEATER WIRE PLUG ON AN ANGLE. FOR BENT OR SPLIT PINS REPORTED TO US BY HEALTHCARE FACILITIES, IT IS IMPOSSIBLE FOR US TO DETERMINE WHETHER THE PINS WERE BENT OR SPLIT DURING PRODUCTION OR BY THE END USER. (B)(4).
A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT THEY WERE UNABLE TO CONNECT THE HEATER WIRE ADAPTOR TO THE INSPIRATORY LIMB OF THREE RT340 BREATHING CIRCUITS BECAUSE THE HEATER WIRE PINS WERE DAMAGED. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT340 | 120404 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |