AU2700 CLINICAL CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2012-01716
- Event Type
- Malfunction
- Date Received
- September 18, 2012
- Date of Event
- April 15, 2011
- Report Date
- July 2, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K003721
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE CUSTOMER INITIALLY EXPERIENCED LOW FLIER IN (B)(6) OF 2011, AND A FIELD SERVICE ENGINEER (FSE) RESOLVED THE ISSUE BY REPLACING THE ION-SELECTIVE ELECTRODE (ISE) D-POT. THE EVENT RE-OCCURRED ONCE IN (B)(6) OF 2012. AT THOSE JUNCTURES SEVERAL HARDWARE PARTS WERE REPLACED, CLEANED, AND EXCHANGED. IN ADDITION, INSTRUMENT PART ALIGNMENTS WERE VERIFIED. CALIBRATIONS WERE VERIFIED AND CUSTOMER QC WAS ACCEPTABLE. A SERVICE REQUEST WAS INITIATED. A FIELD SERVICE ENGINEER (FSE) MONITORED THE SITUATION WITH THE CUSTOMER FOR THE NEXT TWO WEEKS. NO FLYERS WERE NOTED OR REPORTED, HOWEVER, THE CUSTOMER EXPRESSED CONCERNS ABOUT THE STATE OF THE D-POT. THE D-POT STATE WAS MONITORED FOR THE NEXT 6 DAYS WITH NO FLYERS BEING REPORTED. NO DOCUMENTATION WAS PROVIDED AT THE TIME OF EVENT. THE FAILURE MODE OF THE EVENT IS UNKNOWN. THE ISSUE WAS DISCOVERED DURING THE REVIEW OF SERVICE RECORDS. AT THE TIME OF THE SERVICE THE ISSUE WAS RESOLVED BUT THE FIELD PERSONNEL DID NOT INITIATE A RECORD OF THE INCIDENT IN THE DATABASE. BECKMAN COULTER HAS RE-MEDIATED THIS PROCESS AS PART OF THE QUALITY SYSTEM RE-MEDIATION EFFORT AND TRAINING HAS BEEN PROVIDED TO ENSURE THIS GAP IS RECOGNIZED AND ADDRESSED THROUGH A COMPLAINT RECORD. THIS EVENT IS ASSOCIATED WITH MDR REPORTS LISTED BELOW: 2050012-2012-01622 (SUBMITTED ON TIME), 2050012-2012-01714 (SUBMITTED LATE, PAST 30 DAYS), 2050012-2012-01715 (SUBMITTED LATE, PAST 30 DAYS), 2050012-2012-01726 (SUBMITTED LATE, PAST 30 DAYS).
ON (B)(6) 2012, THE CUSTOMER CONTACTED BECKMAN COULTER INC., (BEC) TO REPORT TWO HIGH ISE FLIERS THAT WERE GENERATED ON THE AU2700 CLINICAL CHEMISTRY ANALYZER ON (B)(6) 2011, AS WELL AS A DIRTY D-POT. THE LOW FLIERS WERE 10 MMOL/L BETWEEN TESTING AND WERE NOT CONCURRENT WITH THE AU ANALYZER IN THE LABORATORY. ACTUAL PATIENT DATA/VALUES WERE NOT PROVIDED. THE CUSTOMER REPEATED THE TEST ON THE SAME INSTRUMENT AND AN ALTERNATE ANALYZER. WHEN REPEATED ON AN ALTERNATE INSTRUMENT 7MMOL/L DIFFERENCE IN RESULTS WERE OBTAINED AND WHEN REPEATED ON THE SAME ANALYZER 9 MMOL/L DIFFERENCE WAS OBTAINED. ERRONEOUS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY AND NO CHANGES TO PATIENT TREATMENT ARE ATTRIBUTED TO OR CONNECT TO THIS EVENT. CONTROLS WERE RUN BEFORE AND AFTER THE EVENT AND THE PERFORMANCE WERE ACCEPTABLE. THE UNIT IS CURRENTLY PERFORMING WITHIN QC SPECIFICATIONS WITH RESPECT TO CONTROLS (ACCURACY) AND REPRODUCIBILITY (PRECISION).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AU2700 CLINICAL CHEMISTRY ANALYZER | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER, INC. | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |