FDA Adverse Event Injury Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 2749429 · Received September 18, 2012

Report

Report Number
2531779-2012-10796
Event Type
Injury
Date Received
September 18, 2012
Report Date
August 21, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW UP #1 SUBMITTED: (B)(4) 2012. DEVICE EVALUATION: A RETAINED CARTRIDGE SAMPLE FROM LOT # B201725 WAS EVALUATED. A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A LEAK TEST, FILL TEST, AND FORCE TEST WAS PERFORMED WITH NO FAILURES OBSERVED. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2012, THE PATIENT CLAIMED HER BLOOD GLUCOSE WAS ELEVATED BETWEEN 300-500 MG/DL DUE TO OCCLUSION ISSUE. DURING TROUBLESHOOTING, THE PATIENT NOTED THE CANNULA WAS KINKED AND BENT AFTER EACH OCCLUSION ISSUE OR HIGH BLOOD GLUCOSE EPISODE. IN ADDITION, THE PATIENT WAS NOT USING THE CARTRIDGE PER THE INSTRUCTION FOR USE (IFU). THE CARTRIDGE ALLEGEDLY WAS NOT CYCLE BEFORE FILLING WITH INSULIN. THIS COMPLAINT IS BEING REPORTED DUE TO POSSIBLE USE ERROR AGAINST THE CARTRIDGE AND BECAUSE THE PATIENT HAD ELEVATED BLOOD GLUCOSE BETWEEN 300-500 MG/DL AFTER THE ONSET OF THE OCCLUSION ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM IR1200/1250/2020/OTP CART B201725

Patients

Seq Age Sex Outcome Treatment
1 84 YR Life Threatening