ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2012-10796
- Event Type
- Injury
- Date Received
- September 18, 2012
- Report Date
- August 21, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW UP #1 SUBMITTED: (B)(4) 2012. DEVICE EVALUATION: A RETAINED CARTRIDGE SAMPLE FROM LOT # B201725 WAS EVALUATED. A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A LEAK TEST, FILL TEST, AND FORCE TEST WAS PERFORMED WITH NO FAILURES OBSERVED. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2012, THE PATIENT CLAIMED HER BLOOD GLUCOSE WAS ELEVATED BETWEEN 300-500 MG/DL DUE TO OCCLUSION ISSUE. DURING TROUBLESHOOTING, THE PATIENT NOTED THE CANNULA WAS KINKED AND BENT AFTER EACH OCCLUSION ISSUE OR HIGH BLOOD GLUCOSE EPISODE. IN ADDITION, THE PATIENT WAS NOT USING THE CARTRIDGE PER THE INSTRUCTION FOR USE (IFU). THE CARTRIDGE ALLEGEDLY WAS NOT CYCLE BEFORE FILLING WITH INSULIN. THIS COMPLAINT IS BEING REPORTED DUE TO POSSIBLE USE ERROR AGAINST THE CARTRIDGE AND BECAUSE THE PATIENT HAD ELEVATED BLOOD GLUCOSE BETWEEN 300-500 MG/DL AFTER THE ONSET OF THE OCCLUSION ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM IR1200/1250/2020/OTP CART | B201725 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR | Life Threatening |