FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2742357 · Received September 5, 2012

Report

Report Number
1720753-2012-06918
Event Type
Malfunction
Date Received
September 5, 2012
Date of Event
August 20, 2012
Report Date
September 5, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND CLEANED AND RE-LUBRICATED THE HIGH VOLTAGE CANDLESTICK CONNECTORS. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WAS ARCING AND SHUT DOWN WITHOUT COMMAND. THERE IS NO REPORT OF PATIENT INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1