FDA Adverse Event
Injury
Summary report: N
VASCO TIP BI 9/58
MDR report key: 274230
·
Received April 12, 2000
Report
- Report Number
- 2247185-2000-00003
- Event Type
- Injury
- Date Received
- April 12, 2000
- Date of Event
- March 31, 2000
- Report Date
- April 3, 2000
- Manufacturer
- VASCOR MEDICAL CORP.
- Product Code
- DTB
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Location
- NY, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
SIGNIFICANT DROP IN BIPOLAR IMPEDANCE. FROM 894-1284 1ST YEAR POST IMPLANT, 922-976 OHMS IN 1998, 741-811 OHMS IN 11/99, 753-826 OHMS IN 12/99 AND ON 3/31/2000 316-327 OHMS BIPOLAR AND 486-491 OHMS UNIPOLAR. THRESHOLDS AND CAPTURE CURRENTLY ACCEPTABLE BUT UNDERSENSING IN BIPOLAR OCCURS AT 2.5MV.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VASCO TIP BI 9/58 | PERMANENT PACING LEAD | DTB | VASCOR MEDICAL CORP. | 111-256/58 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 86 YR | Required Intervention | INTERMEDICS #292-05 (FROM 02/07/1997). |