FDA Adverse Event Injury Summary report: N

GUARDIAN

MDR report key: 274207 · Received April 17, 2000

Report

Report Number
MW1018689
Event Type
Injury
Date Received
April 17, 2000
Date of Event
March 30, 2000
Report Date
April 3, 2000
Manufacturer
SUNRISE MEDICAL CCG, INC.
Product Code
FSA
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

TWO CNA'S HOOKED UP THE HOYER LIFT PAD ON TO THE CHAINS, THEN THEY PROCEEDED TO LIFT THE RESIDENT UP IN THE HOYER LIFT. THEY MOVED THE HOYER LIFT AWAY FORM THE BED DUE TO HEIGHT OF MAXI FLO MATTRESS TO RECEIVE ACCURATE WEIGHT. BOTH CNA'S HEARD A POP AND RESIDENT WAS ON THE FLOOR. FACILITY DID AN INSPECTION OF ALL MECHANICAL PARTS AND SLINGS, FOUND IF YOU PLACE THE J HOOK IN A BEND THAT THIS COULD MAKE THE LOAD OF THE ARM TURN LOOSE IF DEFECT IN THE J HOOK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GUARDIAN HOYER LIFT FSA SUNRISE MEDICAL CCG, INC. C 11LA *

Patients

Seq Age Sex Outcome Treatment
1 * Hospitalization