FDA Adverse Event Malfunction Summary report: N

ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT

MDR report key: 2741942 · Received September 12, 2012

Report

Report Number
9611451-2012-00618
Event Type
Malfunction
Date Received
September 12, 2012
Date of Event
August 13, 2012
Report Date
August 14, 2012
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
K983112
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT RT340 BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. IT WAS PRESSURE TESTED AND VISUALLY INSPECTED TO CHECK FOR THE REPORTED LEAK. THE RETURNED BREATHING CIRCUIT WAS SUBSEQUENTLY SUBMERGED IN A WATER BATH. RESULTS: THE PRESSURE TEST REVEALED THAT THE RETURNED BREATHING CIRCUIT EXHIBITED SIGNIFICANT LEAK. VISUAL INSPECTION IDENTIFIED THAT THERE WAS INSUFFICIENT GLUE IN THE PROXIMAL CONNECTOR OF THE EXPIRATORY LIMB. THE WATER BATH TEST HIGHLIGHTED THAT THIS WAS THE SOURCE OF THE REPORTED LEAK. A LOT CHECK DID NOT REVEAL ANY OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120419. CONCLUSION: ALL BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE TESTED PRIOR TO LEAVING THE PRODUCTION LINE. ANY BREATHING CIRCUITS THAT FAIL ARE REJECTED. THIS SUGGESTS THAT THE PROXIMAL CONNECTOR ONLY STARTED TO LEAK ONCE IT WAS SET UP FOR USE. THE USER INSTRUCTIONS SUPPLIED WITH THE RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT STATE THE FOLLOWING: PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT; SET APPROPRIATE VENTILATOR ALARMS. (B)(4)..

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE THAT AN RT340 ADULT BREATHING CIRCUIT FAILED THE LEAK TEST ON A SERVO-I-VENTILATOR. THEY FURTHER STATED THAT THE EXPIRATORY LIMB HAD AN AIR LEAK BUT NO PIN HOLE WAS FOUND. THIS WAS FOUND PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LTD RT340 120419

Patients

Seq Age Sex Outcome Treatment
1 SERVO-I-VENTIALTOR