ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2012-00618
- Event Type
- Malfunction
- Date Received
- September 12, 2012
- Date of Event
- August 13, 2012
- Report Date
- August 14, 2012
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- K983112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT RT340 BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. IT WAS PRESSURE TESTED AND VISUALLY INSPECTED TO CHECK FOR THE REPORTED LEAK. THE RETURNED BREATHING CIRCUIT WAS SUBSEQUENTLY SUBMERGED IN A WATER BATH. RESULTS: THE PRESSURE TEST REVEALED THAT THE RETURNED BREATHING CIRCUIT EXHIBITED SIGNIFICANT LEAK. VISUAL INSPECTION IDENTIFIED THAT THERE WAS INSUFFICIENT GLUE IN THE PROXIMAL CONNECTOR OF THE EXPIRATORY LIMB. THE WATER BATH TEST HIGHLIGHTED THAT THIS WAS THE SOURCE OF THE REPORTED LEAK. A LOT CHECK DID NOT REVEAL ANY OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120419. CONCLUSION: ALL BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE TESTED PRIOR TO LEAVING THE PRODUCTION LINE. ANY BREATHING CIRCUITS THAT FAIL ARE REJECTED. THIS SUGGESTS THAT THE PROXIMAL CONNECTOR ONLY STARTED TO LEAK ONCE IT WAS SET UP FOR USE. THE USER INSTRUCTIONS SUPPLIED WITH THE RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT STATE THE FOLLOWING: PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT; SET APPROPRIATE VENTILATOR ALARMS. (B)(4)..
A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE THAT AN RT340 ADULT BREATHING CIRCUIT FAILED THE LEAK TEST ON A SERVO-I-VENTILATOR. THEY FURTHER STATED THAT THE EXPIRATORY LIMB HAD AN AIR LEAK BUT NO PIN HOLE WAS FOUND. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT340 | 120419 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | SERVO-I-VENTIALTOR |