FREESTYLE FREEDOM
Report
- Report Number
- 2954323-2012-06559
- Event Type
- Malfunction
- Date Received
- September 11, 2012
- Date of Event
- August 28, 2012
- Report Date
- August 28, 2012
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER RETURNED METER SERIAL NUMBER ((B)(4)). THE RETURNED METER WAS INVESTIGATED WITH RETAINED STRIPS. THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE READINGS THE CUSTOMER REPORTED WERE NOT FOUND IN METER MEMORY.
REQUESTED PRODUCT WAS NOT RECEIVED FOR INVESTIGATION. RETAINED TEST STRIP SAMPLES FROM THE SAME LOT REPORTED BY THE CUSTOMER (1260946) WERE TESTED WITH CONTROL SOLUTION INSTEAD. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED. THE COMPLAINT IS NOT CONFIRMED.
THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED. (B)(4).
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 77 MG/DL 283 MG/DL, 156 MG/DL AND 45 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE FREEDOM | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 1260946 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR |