FDA Adverse Event Death Summary report: N

QUADROX-ID

MDR report key: 2732985 · Received September 4, 2012

Report

Report Number
8010762-2012-00033
Event Type
Death
Date Received
September 4, 2012
Date of Event
July 23, 2012
Report Date
August 9, 2012
Manufacturer
MAQUET CARDIOPULMONARY AG
Product Code
DTZ
PMA / PMN Number
K112360
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

MAQUET MEDICAL SYSTEMS, USA SUBMITS THIS REPORT ON BEHALF OF THE LEGAL MFR OF THE DEVICE MAQUET (B)(4). MAQUET (B)(4) PROVIDES PRODUCT FAILURE INVESTIGATION, ANALYSIS AND RESOLUTION FOR THE DEVICE DESCRIBED IN THIS REPORT. THE FACILITY MEDWATCH STATE LOT #70075847, HOWEVER, DURING F/U WITH THE FACILITY IT WAS NOTED THAT TWO OXYGENATORS WERE USED BUT THEY WERE UNABLE TO DETERMINE WHICH ONE WAS INVOLVED IN THE INCIDENT. ADDITIONALLY, THE LOT #'S USED WERE 70072248 AND 70079206 AND NOT LOT #70075847.

Description of Event or Problem · 1

EVENT DESCRIPTION FROM THE HOSP - "THE PT WAS IN THE CARDIAC INTENSIVE CARE UNIT (CICU), ON EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO). THE MEDICAL TEAM HAD FINISHED THE SECOND EVAL FOR BRAIN DEATH WHICH WAS CONSISTENT WITH BRAIN DEATH. THEY WERE AWAITING RESULTS OF A PRIOR EEG. A PHYSICIAN LIFTED AND LOWERED THE HEAD OF THE PT'S BED. THIS APPEARS TO HAVE CAUSED A PART OF THE MAQUET POST OXYGENATOR. CONNECTION OUTLET TO BREAK IN TWO, CAUSING A DISRUPTION IN THE ECMO CIRCUIT WITH RAPID BLOOD LOSS. ONE OF THE PHYSICIANS RE-ATTACHED THE PARTS LEAVING ONLY A SMALL DRIPPING OF BLOOD. DURING THIS EVENT, IT WAS LEARNED THAT THE PRIOR EEG CONFIRMED BRAIN DEATH AND THE PT WAS PRONOUNCED DEAD". F/U WITH THE FACILITY ON (B)(4) 2012: THIS WAS NOT A FAILURE OF THE OXYGENATOR. WHEN THE HEAD OF THE BED WAS BEING LOWERED ONE OF THE ARM RAILS HIT THE OUTLET AND BROKE IT OFF. THE PT HAD BEEN ON SUPPORT SINCE (B)(6). (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUADROX-ID OXYGENATOR DTZ MAQUET CARDIOPULMONARY AG BEQ-HMOD70000-USA NI

Patients

Seq Age Sex Outcome Treatment
1 48 YR Death