FDA Adverse Event Other Summary report: N

L-70 NORMOTHERMIC FLUID WARMING SET

MDR report key: 272318 · Received March 28, 2000

Report

Report Number
1221261-2000-00001
Event Type
Other
Date Received
March 28, 2000
Date of Event
February 21, 2000
Report Date
March 20, 2000
Manufacturer
SIMS LEVEL 1, INC.
Product Code
LGZ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

SIMS LEVEL 1 RECEIVED A REPORT FROM SIMS CANADA, LTD THAT ABOUT HALF HOUR-HOUR INTO A SURGICAL PROCEDURE, AN ANETHESIA PHYSICIAN NOTICED THAT THE DRIP CHAMBER OF THE Y SET WAS COMPLETELY FILLED WITH FLUID AND PT WAS IN PRONE POSITION AND THE ARM WAS BENDING FROM TOO MUCH FLUID BEING INFUSED AT A FAST RATE. THE PHYSICIAN THEN DETERMINED THAT THE FLUID FROM THE WATER RESEVOIR HAD CROSSED OVER INTO THE IV PATH. THE DISPOSABLE SET WAS RETURNED TO LEVEL 1 AND UPON VISUAL EXAMINATION, IT WAS NOTED THAT THERE WAS A NEEDLE-STICK PUNCTURE IN THE TOP IF THE Y-INJECTION PORT. UPON DESTRUCTIVE TESTING AND VISUALIZATION UNDER A MICROSCOPE, IT WAS NOTED THAT THERE WAS A PUNCTURE FROM A NEEDLE, THROUGH THE CONNECTOR WALL AND INTO THE IV PATH OF THE TUBING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 L-70 NORMOTHERMIC FLUID WARMING SET IV AND BLOOD ADMINSTRATION SET LGZ SIMS LEVEL 1, INC. L-70 199811A

Patients

Seq Age Sex Outcome Treatment
1 NO INFO Other