FDA Adverse Event
Malfunction
Summary report: N
CORE MICRO DRILL
MDR report key: 2722101
·
Received August 24, 2012
Report
- Report Number
- 1811755-2012-03225
- Event Type
- Malfunction
- Date Received
- August 24, 2012
- Date of Event
- July 26, 2012
- Report Date
- July 26, 2012
- Manufacturer
- STRYKER INSTRUMENTS, INSTRUMENTS DIV.
- Product Code
- HBE
- PMA / PMN Number
- K032303
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE DRILL WAS RECEIVED BY THE MFR, HOWEVER THE OVERHEATING COMPLAINT COULD NOT BE REPLICATED. EVAL OF INTERNAL COMPONENTS REVEALED THAT THE ROTOR WAS LODGED IN THE MOTOR ASSEMBLY DUE TO CORROSION CONTAMINATION. THIS CONDITION CAN RESULT IN EXCESSIVE FRICTION AND HEAT GENERATION.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING AN EVAL CONDUCTED BY A MFR FIELD SERVICE TECH, THE DRILL HEATED UP. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PT INVOLVEMENT. NO USER INJURY WAS REPORTED AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.