FDA Adverse Event Malfunction Summary report: N

INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT

MDR report key: 2721996 · Received August 30, 2012

Report

Report Number
9611451-2012-00589
Event Type
Malfunction
Date Received
August 30, 2012
Report Date
August 3, 2012
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K034026. METHOD: THE COMPLAINT BREATHING CIRCUIT WAS RECEIVED AT FISHER & PAYKEL HEALTHCARE (B)(4) AND WAS VISUALLY INSPECTED. RESULTS: THE DRYLINE WAS FOUND TO BE MISSING FROM THE BREATHING CIRCUIT KIT. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120526. CONCLUSION: EACH BREATHING CIRCUIT KIT CONSISTS OF A NUMBER OF COMPONENTS GROUPED TOGETHER DURING PRODUCTION. IT IS LIKELY THAT OPERATOR ERROR HAS RESULTED IN THE OMITTED DRYLINE. THERE ARE STANDARD OPERATING PROCEDURES (SOPS) IN PLACE TO ASSIST OPERATORS ON THE PRODUCTION LINE CORRECTLY PACK BREATHING CIRCUITS. THIS CONSISTS OF A SPECIFIC PACK CARD, DETAILING ALL COMPONENTS REQUIRED WHICH MUST BE DISPLAYED AT THE PACKING STATION. (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT WHEN THEY OPENED AN RT265 INFANT BREATHING CIRCUIT KIT THEY FOUND THAT THE DRYLINE TUBE WAS MISSING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LTD RT265 120526

Patients

Seq Age Sex Outcome Treatment
1