FDA Adverse Event Malfunction Summary report: N

COMPOUNDERS

MDR report key: 2716642 · Received August 28, 2012

Report

Report Number
6000001-2012-13519
Event Type
Malfunction
Date Received
August 28, 2012
Date of Event
August 8, 2012
Report Date
August 8, 2012
Manufacturer
BAXTER HEALTHCARE - BUFFALO GROVE
Product Code
NEP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS IS A CLASS II 510(K) EXEMPT DEVICE WITH THE ADDITIONAL 510(K) OF K955622. THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE CAUSE WAS DETERMINED TO BE A DAMAGED UMBILICAL CABLE. TO CORRECT THE CONDITION, THE UMBILICAL CABLE WAS REPLACED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, AN AUTOMIX COMPOUNDER REQUIRED THE REPLACEMENT OF A DAMAGED UMBILICAL CABLE. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THIS WAS NOT REPORTED BY THE CUSTOMER. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COMPOUNDERS SYSTEM/DEVICE, PHARMACY COMPOUNDING NEP BAXTER HEALTHCARE - BUFFALO GROVE

Patients

Seq Age Sex Outcome Treatment
1