UNICEL DXC 800 SYNCHRON SYSTEM
MDR report key: 2716039
·
Received August 27, 2012
Report
- Report Number
- 2050012-2012-01430
- Event Type
- Malfunction
- Date Received
- August 27, 2012
- Date of Event
- August 2, 2012
- Report Date
- August 2, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
CUSTOMER REPORTED TO BECKMAN COULTER, INC. (BEC) THAT THE BLADE WASH ON THE CAP PIERCER OF THE UNICEL DXC 800 SYNCHRON SYSTEM WAS OVERFLOWING. CUSTOMER REPORTED THAT ERRONEOUS RESULTS WERE NOT GENERATED. THERE WAS NO REPORT OF ANY ADVERSE EVENT OR INJURY REQUIRING MEDICAL INTERVENTION OR PATIENT TREATMENT. BEC FIELD SERVICE ENGINEER (FSE) FOUND DEBRIS OCCLUDING THE QUICK DISCONNECT FITTING ON LINE 180. THE FSE REPLACED THE FITTING.