FDA Adverse Event Malfunction Summary report: N

UNICEL DXC 800 SYNCHRON SYSTEM

MDR report key: 2716039 · Received August 27, 2012

Report

Report Number
2050012-2012-01430
Event Type
Malfunction
Date Received
August 27, 2012
Date of Event
August 2, 2012
Report Date
August 2, 2012
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED TO BECKMAN COULTER, INC. (BEC) THAT THE BLADE WASH ON THE CAP PIERCER OF THE UNICEL DXC 800 SYNCHRON SYSTEM WAS OVERFLOWING. CUSTOMER REPORTED THAT ERRONEOUS RESULTS WERE NOT GENERATED. THERE WAS NO REPORT OF ANY ADVERSE EVENT OR INJURY REQUIRING MEDICAL INTERVENTION OR PATIENT TREATMENT. BEC FIELD SERVICE ENGINEER (FSE) FOUND DEBRIS OCCLUDING THE QUICK DISCONNECT FITTING ON LINE 180. THE FSE REPLACED THE FITTING.