FDA Adverse Event
Injury
Summary report: N
TOSHIBA
MDR report key: 2714870
·
Received February 3, 2012
Report
- Report Number
- 2020563-2012-00003
- Event Type
- Injury
- Date Received
- February 3, 2012
- Date of Event
- December 16, 2011
- Report Date
- January 5, 2012
- Manufacturer
- TOSHIBA MEDICAL SYSTEMS CORPORATION
- Product Code
- JAK
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- RADIOLOGIC TECHNOLOGIST
Narratives
Additional Manufacturer Narrative · 1
THE PT WAS ON THE COUCH HEAD FIRST AND SUPINE WITH ARMS DOWN BY THE SIDE AND HANDS ON CHEST. AS THE PT WAS BEING SENT INTO THE GANTRY, THE ARM WAS PINCHED BY THE MOVING COUCH TOP BETWEEN THE COUCH TOP AND COUCH. THE VINYL COVERING FOR THE COUCH TOP THAT SLIPS OVER THE COUCH TOP PAD DID NOT COVER THE GAP BETWEEN THE COUCH TOP AND THE COUCH. THERE WAS A SHEET UNDER THE PT, BUT IT WAS NOT COVERING THE ENTIRE COUCH TOP. THE PT WAS ONLY PINCHED. THERE WAS NO BREAKING OF THE SKIN OR NEED FOR MEDICAL TREATMENT.
Description of Event or Problem · 1
AS A PT WAS BEING SENT INTO THE GANTRY, THE PT'S ARM WAS PINCHED BY THE MOVING COUCH TOP BETWEEN THE COUCH TOP AND THE COUCH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOSHIBA | COMPUTER TOMOGRAPHY SYSTEM | JAK | TOSHIBA MEDICAL SYSTEMS CORPORATION | TSX-101A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |