COSYCOT INFANT WARMER
Report
- Report Number
- 9611451-2012-00562
- Event Type
- Malfunction
- Date Received
- August 20, 2012
- Report Date
- July 27, 2012
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- FMT
- PMA / PMN Number
- K971695
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4). SERIAL NUMBERS AND DEVICE MANUFACTURER DATES: 090224000308 - 24 FEBRUARY 2009; 081030001612 - 30 OCTOBER 2008; 090224000300 - 24 FEBRUARY 2009; 090224000306 - 24 FEBRUARY 2009; 090224000311 - 24 FEBRUARY 2009; 090224000305 - 24 FEBRUARY 2009; 030204000136 - 04 FEBRUARY 2003; 030129000120 - 29 JANUARY 2003; 030129000121 - 29 JANUARY 2003. THE COMPLAINT DEVICES ARE NOT EXPECTED TO BE RETURNED TO FISHER & PAYKEL HEALTHCARE (FPH). THREE PHOTOGRAPHS OF THE COMPLAINT HEATER HEADS WERE PROVIDED TO US BY THE HOSPITAL. OUR INVESTIGATION IS BASED ON THE INFORMATION PROVIDED BY THE HOSPITAL. THE PHOTOGRAPHS REVEALED CRACKING AND FLAKING OF THE PLASTIC ON THE UPPER HEATER HEAD CASINGS. CRAZING WAS ALSO OBSERVED ON THE DOME AREA OF THE UPPER HEAD. THE HOSPITAL REPORTED THAT "(B)(6)" MULTI SURFACE DETERGENT WIPES WERE USED TO CLEAN THE COMPLAINT DEVICES. A LOT CHECK REVEALED ONE OTHER COMPLAINT OF THIS TYPE FOR LOT NUMBER 090224, AND NO OTHER COMPLAINT OF THIS TYPE FOR LOT NUMBERS 081030, 030204, AND 030129. CONCLUSION: THE CLEANING SOLUTION USED BY THE HOSPITAL, "CLINITEX", IS NOT A FPH RECOMMENDED CLEANING SOLUTION. THE CRACKING AND CRAZING OF THE COMPLAINT WARMER HEADS ARE MOST LIKELY RELATED TO A KNOWN PROBLEM OF INCOMPATIBLE CLEANING SOLUTIONS REACTING WITH THE POLYCARBONATE PLASTIC AND ACRYLIC COMPONENTS OF THE INFANT WARMER. WHEN THE HEATER HEAD BEGINS TO WARM UP DURING USE A CHEMICAL REACTION WITH THE INCOMPATIBLE CLEANING SOLUTION RESULTS IN GRADUAL CRACKING AND CRAZING OF THE HEATER HEAD. THE INFANT WARMER SERVICE MANUAL FOR THE AFFECTED UNIT FEATURES A DIAGRAM OF THE INFANT WARMER WHICH HIGHLIGHTS POLYCARBONATE AND ACRYLIC COMPONENTS AND STATES THE FOLLOWING: CAUTION DO NOT CLEAN THE HIGHLIGHTED PLASTIC SURFACES WITH PROPRIETARY CLEANING PRODUCTS CONTAINING EITHER HYDROXIDES, HYPOCHLORITES, PEROXIDES, GLUTARALDEHYDE OR CLEANING PRODUCTS WITH A GREATER THAN 30% ALCOHOL BASE; CAUTION THE CHEMICALS USED IN THESE PROPRIETARY CLEANING PRODUCTS MAY LEAD TO DISCOLOURATION, CRAZING AND EVENTUAL CRACKING OF THE HIGHLIGHTED PLASTIC SURFACES. AS PART OF CONTINUOUS IMPROVEMENT ON 2 JUNE 2010, WE CHANGED THE MATERIAL USED IN THE MANUFACTURE OF THE HEAD HOUSING TO ONE THAT OFFERS GREATER RESISTANCE TO ENVIRONMENTAL STRESS CRACKING. WE ALSO REVISED OUR CLEANING INSTRUCTIONS TO RECOMMEND SPECIFIC PROPRIETARY CLEANING WIPES.
A HOSPITAL IN (B)(6) REPORTED THAT THE HEATER HEADS OF 9 IW952 INFANT WARMERS WERE CRACKED. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COSYCOT INFANT WARMER | FMT | FMT | FISHER & PAYKEL HEALTHCARE LTD | IW952 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |