FDA Adverse Event Malfunction Summary report: N

CHEMICLAVE

MDR report key: 27061 · Received October 6, 1995

Report

Report Number
MW4000832
Event Type
Malfunction
Date Received
October 6, 1995
Date of Event
March 10, 1995
Report Date
March 20, 1995
Manufacturer
MDT BIO/MDT DIV.
Product Code
MLR
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

ON 3/10/95 A DENTAL TECH NOTICED SMOKE COMING FROM THE UNIT. FIRE DEPT WAS CALLED & RESPONDED. WHILE AWAITING THE ARRIVAL OF THE FIRE DEPT THE UNIT WAS UNPLUGGED & THE SMOKE DISSIPATED. AN INVESTIGATION BY THE LOCAL BIO-MEDICAL EQUIPMENT REPAIR WORK CENTER REVEALED THAT THE SMOKE WAS CAUSED BY A SOLID STATE RELAY ON THE I/O PRINTED CIRCUIT BOARD THAT OVERHEATED & CAUGHT FIRE. THE UNIT'S CIRCUIT BREAKER & FUSES DID NOT TRIP OR BLOW.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CHEMICLAVE CHEMICLAVE MLR MDT BIO/MDT DIV. 8000D

Patients

Seq Age Sex Outcome Treatment
1 *