FDA Adverse Event
Malfunction
Summary report: N
CHEMICLAVE
MDR report key: 27061
·
Received October 6, 1995
Report
- Report Number
- MW4000832
- Event Type
- Malfunction
- Date Received
- October 6, 1995
- Date of Event
- March 10, 1995
- Report Date
- March 20, 1995
- Manufacturer
- MDT BIO/MDT DIV.
- Product Code
- MLR
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
ON 3/10/95 A DENTAL TECH NOTICED SMOKE COMING FROM THE UNIT. FIRE DEPT WAS CALLED & RESPONDED. WHILE AWAITING THE ARRIVAL OF THE FIRE DEPT THE UNIT WAS UNPLUGGED & THE SMOKE DISSIPATED. AN INVESTIGATION BY THE LOCAL BIO-MEDICAL EQUIPMENT REPAIR WORK CENTER REVEALED THAT THE SMOKE WAS CAUSED BY A SOLID STATE RELAY ON THE I/O PRINTED CIRCUIT BOARD THAT OVERHEATED & CAUGHT FIRE. THE UNIT'S CIRCUIT BREAKER & FUSES DID NOT TRIP OR BLOW.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CHEMICLAVE | CHEMICLAVE | MLR | MDT BIO/MDT DIV. | 8000D |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |