ADULT INSPIRATORY HEATED BREATHING CIRCUIT
Report
- Report Number
- 9611451-2012-00559
- Event Type
- Malfunction
- Date Received
- August 16, 2012
- Date of Event
- July 16, 2012
- Report Date
- July 27, 2012
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE RT206 IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. METHOD: ONLY THE DRYLINE TUBE OF THE COMPLAINT DEVICE WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR VISUAL INSPECTION. RESULTS: VISUAL INSPECTION REVEALED THAT A HOLE WAS FOUND APPROXIMATELY 2 CM AWAY FROM THE CONNECTOR. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS TYPE FOR LOT NUMBER 120126. CONCLUSION: IT WAS IDENTIFIED THAT HOLE DAMAGE TO THE DRYLINE TUBE COULD HAVE BEEN CAUSED DURING THE MANUFACTURING PROCESS, ALTHOUGH IT IS NOT A BATCH RELATED ISSUE. DURING THE PRODUCTION OF THE DRYLINE TUBES, A PRESSURE TEST IS PERFORMED EVERY 10 MINUTES AND THOSE THAT FAIL ARE SET ASIDE FOR FURTHER INSPECTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT206 BREATHING CIRCUIT STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS." (B)(4).
A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT AN RT206 ADULT BREATHING CIRCUIT FAILED THE VENTILATOR LEAK TEST AND A PINHOLE WAS FOUND ON THE DRYLINE. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADULT INSPIRATORY HEATED BREATHING CIRCUIT | BTT | BTT | FISHER & PAYKEL HEALTHCARE LTD | RT206 | 120126 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |