FDA Adverse Event
Malfunction
Summary report: N
ADVANTA BED
MDR report key: 2701594
·
Received August 10, 2012
Report
- Report Number
- 1824206-2012-04747
- Event Type
- Malfunction
- Date Received
- August 10, 2012
- Date of Event
- July 25, 2012
- Report Date
- July 25, 2012
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECH REPLACED THE CIB CIRCUIT BOARD TO REPAIR THE BED.
Description of Event or Problem · 1
INFO RECEIVED INDICATES THE CIB CIRCUIT BOARD HAS NO POWER BECAUSE NONE OF THE LEDS ARE ILLUMINATED. WHEN THE TECH OPENED THE CIB ASSEMBLY, HE FOUND FLUID ON THE CIB CIRCUIT BOARD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADVANTA BED | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1600 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |