FDA Adverse Event Malfunction Summary report: N

SPECTRUM

MDR report key: 2699380 · Received August 8, 2012

Report

Report Number
2699380
Event Type
Malfunction
Date Received
August 8, 2012
Date of Event
August 8, 2012
Report Date
August 8, 2012
Manufacturer
SIGMA INTERNATIONAL
Product Code
FRN
Product Problem
Yes
Report Source
User Facility report
Reporter Location
SD, US

Narratives

Description of Event or Problem · 1

WHILE IN PROCESS OF INVESTIGATING AN ISSUE WITH ONE OF THESE PUMPS, IT CAME TO OUR ATTENTION THAT THEY OPERATE BEHIND THE SCENES ON GMT GREENWICH MEAN TIME AND THE INTERNAL CLOCK CAN NOT BE RESET TO OUR LOCAL TIME. THUS, THE ALARM RECORD MAY BE ACCURATE TO THE MINUTE, BUT IT IS NOT ACCURATE TO THE HOUR AND MAKES INVESTIGATING THE SITUATION OF WHETHER IT ALARMED WHEN IT SHOULD HAVE FOR A PROBLEM WITH THE IV VERY DIFFICULT TO FIND. THE INTERNAL CLOCK SHOULD BE ADJUSTABLE TO OUR LOCAL TIME, SO IF AN IV DRIP IS OCCLUDED, THE UPSTREAM OCCLUSION ALARM RECORD SHOWS LOCAL TIME NOT GMT.======================MANUFACTURER RESPONSE FOR IV SMART PUMPS, SIGMA SPECTRUM IV PUMP (PER SITE REPORTER).======================WHEN BIO MED QUESTIONED WHY OUR ATTEMPTS TO RE-SET THE TIME TO CENTRAL TIME WOULD NOT STAY, THEY WERE TOLD THAT THE INTERNAL CLOCKS ARE SET TO GMT ON ALL THE SIGMA SPECTRUM SMART PUMPS FOR STANDARDIZATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPECTRUM INFUSION PUMP FRN SIGMA INTERNATIONAL * *

Patients

Seq Age Sex Outcome Treatment
1 *