FDA Adverse Event Other Summary report: N

MICRO-PACE

MDR report key: 2699180 · Received August 13, 2012

Report

Report Number
1218087-2012-00045
Event Type
Other
Date Received
August 13, 2012
Manufacturer
PACE MEDICAL, INC
Product Code
JOQ
PMA / PMN Number
P920032
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CUSTOMER RETURNED THE DEVICE WITH THE COMPLAINT THAT THE ATRIAL CHANNEL WAS NOT WORKING. THE DEVICE WAS EXAMINED AND FOUND THAT A TRACE NEEDED REPAIR. THE REPAIR WAS MADE. THE DEVICE WAS RE-CALIBRATED AND FINAL TESTED. ALL TEST WERE PASSED AND WAS RETURNED TO THE CUSTOMER. REASON FOR COMPLAINT: POSSIBLE ROUGH HANDLING OR A SUDDEN IMPACT.

Description of Event or Problem · 1

PACE MEDICAL WAS NOTIFIED ON 08/16/2011 BY (B)(6) VIA LETTER THAT UNIT HAD A FAULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICRO-PACE TEMPORARY CARDIAC PACEMAKER JOQ PACE MEDICAL, INC 4570 NA

Patients

Seq Age Sex Outcome Treatment
1