ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2012-00554
- Event Type
- Malfunction
- Date Received
- August 13, 2012
- Date of Event
- July 16, 2012
- Report Date
- July 19, 2012
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BZE
- PMA / PMN Number
- K983112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). LOT NUMBER: 101027,120210, 120320, 120326, MANUFACTURING DATE: 10/27/2010, 02/10/2012, 03/20/2012, 03/26/2012, QUANTITY AFFECTED: (B)(4). METHOD: THE 12 COMPLAINT RT240 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUITS WERE RETURNED TO FISHER & PAYKEL HEALTHCARE (FPH) IN (B)(4) FOR INSPECTION. THE RETURNED DEVICES WERE VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED THAT THE 12 RETURNED BREATHING CIRCUITS EACH HAD A HOLE ON THE DRYLINE TUBE. EACH HOLE WAS LOCATED EITHER 4.5 CM OR 10 CM FROM THE DRYLINE CONNECTOR. A LOT CHECK REVEALED (B)(4) OTHER COMPLAINT FOR LOT NUMBER 101027. A LOT CHECK REVEALED (B)(4) OTHER COMPLAINTS FOR LOT NUMBER 120210. A LOT CHECK REVEALED (B)(4) OTHER COMPLAINTS FOR LOT NUMBER 120320. A LOT CHECK REVEALED (B)(4) OTHER COMPLAINT FOR LOT NUMBER 120326. CONCLUSION: IT WAS IDENTIFIED THAT DAMAGE TO THE RT240 DRYLINE TUBE COULD HAVE BEEN CAUSED DURING THE MANUFACTURING PROCESS. DURING THE PRODUCTION OF THE DRYLINE TUBES, A PRESSURE TEST IS PERFORMED EVERY 10 MINUTES AND THOSE THAT FAIL ARE SET ASIDE FOR FURTHER INSPECTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT240 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "VERIFY THAT VENTILATOR ALARMS ARE SET TO DETECT LOSS OF PRESSURE AND OVER PRESSURE." (B)(4).
A HEALTHCARE FACILITY IN (B)(6) REPORTED TO OUR DISTRIBUTOR THAT 12 UNITS OF RT240 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUITS "ARE NOT PASSING THE LEAK TEST". NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LIMITED | RT240 | SEE H10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |