CELL-DYN EMERALD
Report
- Report Number
- 2919069-2012-00056
- Event Type
- Malfunction
- Date Received
- August 13, 2012
- Date of Event
- August 7, 2012
- Report Date
- August 8, 2012
- Manufacturer
- ABBOTT DIAGNOSTICS DIVISION
- Product Code
- GKZ
- PMA / PMN Number
- K081495
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
AN ABBOTT FIELD SERVICE REPRESENTATIVE (FSR) REPLACED THE RBC COUNT HEAD ON THE CUSTOMER'S CELL-DYN EMERALD ANALYZER TO RESOLVE THE REPORTED ISSUE (THE COUNT HEAD WAS FOUND TO BE WORN FROM NORMAL USE). THE FSR BLEACHED CLEANED AND FLUSHED THE LINES, AND CHECKED THE AUTOCLEAN BOX. THE INSTRUMENT WAS PERFORMING WITHIN LABELING SPECIFICATIONS AND SUBSEQUENT INSTRUMENT OPERATIONS WERE ACCEPTABLE. A REVIEW OF COMPLAINT TRACKING AND TRENDING METRICS WAS PERFORMED AND IDENTIFIED NO ADVERSE TRENDS IN CONJUNCTION WITH THE COMPLAINT ISSUE CURRENTLY UNDER EVALUATION. THE CELL-DYN EMERALD SYSTEM OPERATOR MANUAL, LIST NUMBER 09H40-01,REVISION E, CONTAINS INFORMATION TO ADDRESS THE CUSTOMER'S CURRENT ISSUE. BASED ON THE CURRENT EVALUATION AND INFORMATION RECEIVED, NO PRODUCT MALFUNCTION WAS IDENTIFIED WITH THE CELL-DYN EMERALD INSTRUMENT.
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. (B)(4).
THE CUSTOMER REPORTS ELEVATED PLATELET RESULTS FOR ONE PATIENT SAMPLE GENERATED ON A CELL-DYN EMERALD ANALYZER. AN INITIAL RESULT OF 1217 K/UL RETESTED AT 367 K/UL. THE SAMPLE WAS THEN TESTED ON ANOTHER CELL-DYN EMERALD ANALYSER IN THE LAB AND GENERATED A PLATELET RESULT OF 280 K/UL. CONTROLS HAVE REMAINED WITHIN SPECIFICATIONS. THE CUSTOMER WOULD NOT PROVIDE ANY FURTHER INFORMATION OTHER THAN TO VERIFY THAT THERE IS NO IMPACT TO PATIENT MANAGEMENT REPORTED. THE CUSTOMER REQUESTED A SERVICE CALL TO RESOLVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CELL-DYN EMERALD | GKZ | ABBOTT DIAGNOSTICS DIVISION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CD EMERALD LYSE: LN: 09H47-02 LOT:5212| CD EMERALD DILUENT: LN: 09H48-02 LOT: 5291| CD EMERALD CLEANER: LN: 09H46-02 LOT:5294| CD EMERALD DILUENT: LN: 09H48-02 LOT: 5291| CD EMERALD CLEANER: LN: 09H46-02 LOT:5294| RBC COUNTING HEAD: LN: 8701687701 |