ACCESS 2 IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2012-01672
- Event Type
- Malfunction
- Date Received
- August 9, 2012
- Date of Event
- July 5, 2012
- Report Date
- July 19, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- KXT
- PMA / PMN Number
- K922823/A007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
SERVICE WAS DISPATCHED TO THE SITE FOR THIS EVENT THE FIELD SERVICE ENGINEER (FSE) REPLACED THE TRANSDUCER AS "SUSPECT TRANSDUCER ISSUES" WERE NOTED DURING PREVIOUS FSE VISITS. QUALITY CONTROL (QC) WAS PERFORMED FOLLOWING REPLACEMENT OF THE TRANSDUCER AND RESULTS RECOVERED WITHIN THE LABORATORY'S ESTABLISHED RANGES AND CLOSER TO THE MEAN VALUES ASSIGNED TO EACH LEVEL OF QC. A HARDWARE MALFUNCTION IS THE LIKELY CAUSE OF THIS EVENT
THE CUSTOMER REPORTED THAT AN ERRONEOUS ELEVATED DIGOXIN RESULT ABOVE THE THERAPEUTIC RANGE OF THE ASSAY WAS GENERATED FROM AN ACCESS 2 IMMUNOASSAY SYSTEM FOR ONE PATIENT'S SAMPLE. BECKMAN COULTER INC. ASSESSMENT OF CUSTOMER SUPPLIED DATA INDICATED THAT UPON REPEAT TESTING OF THE SAMPLE ON AN ALTERNATE INSTRUMENT, A LOWER DIGOXIN RESULT WAS GENERATED WHICH WAS CONSIDERED VALID. THE ERRONEOUS DIGOXIN RESULT WAS REPORTED OUTSIDE OF THE LABORATORY; HOWEVER, THERE WERE NO REPORTS OF DEATH, SERIOUS INJURY OR CHANGE TO PATIENT MANAGEMENT ASSOCIATED WITH THIS EVENT. BECKMAN COULTER INC. ASSESSMENT OF CUSTOMER SUPPLIED INSTRUMENT/ASSAY PERFORMANCE DATA, INDICATED THAT LEVEL TWO AND THREE QUALITY CONTROL (QC) RESULTS WERE WITHIN ONE TO TWO STANDARD DEVIATIONS OF THE LABORATORY'S ESTABLISHED RANGE PRIOR TO ANALYSIS OF THE SAMPLE, BUT WERE OUTSIDE THE LABORATORY'S ESTABLISHED RANGE FOLLOWING ANALYSIS OF THE INVOLVED SAMPLE. LEVEL ONE QC RESULTS WERE WITHIN ONE TO TWO STANDARD DEVIATIONS OF THE LABORATORY'S ESTABLISHED RANGE ON THE DATE OF EVENT. CALIBRATIONS PERFORMED PRIOR TO THE EVENT GENERATED PASSING RESULTS WITH ACCEPTABLE PERCENT COEFFICIENTS OF VARIATION. ROUTINE SYSTEM CHECKS PERFORMED PRIOR TO THE EVENT PASSED WITHIN INSTRUMENT SPECIFICATIONS. THE REAGENT AND CALIBRATOR LOTS ASSOCIATED WITH THIS EVENT WERE 1220507 AND 221479 RESPECTIVELY. NO SAMPLE COLLECTION/HANDLING INFORMATION WAS PROVIDED BY THE CUSTOMER
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCESS 2 IMMUNOASSAY SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | KXT | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR |