SYNCHRON LXI 725 SYSTEM
Report
- Report Number
- 2122870-2012-01643
- Event Type
- Malfunction
- Date Received
- August 9, 2012
- Date of Event
- July 16, 2012
- Report Date
- July 16, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- MMI
- PMA / PMN Number
- K023049
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
SERVICE WAS DISPATCHED TO THE SITE FOR THIS EVENT. THE FIELD SERVICE ENGINEER (FSE) FOUND DRIED WASH BUFFER IN THE INCUBATOR TRACK AND THE WASH WHEEL OF THE INSTRUMENT. THE FSE PERFORMED PREVENTIVE MAINTENANCE ON THE INSTRUMENT. AFTER COMPLETION OF THE NECESSARY AND VERIFIED ACTIVITIES THE INSTRUMENT WAS RETURNED INTO OPERATION. A DEFINITIVE ROOT CAUSE HAS NOT BEEN DETERMINED TO DATE FOR THIS EVENT.
THE CUSTOMER REPORTED THAT AN ELEVATED CARDIAC TROPONIN (ACCUTNI) RESULT, WITHIN THE RISK STRATIFICATION RANGE OF THE ASSAY, WAS GENERATED ON AN SYNCHRON LXI 725 SYSTEM FOR A SINGLE PATIENT SAMPLE. THE PATIENT SAMPLE WAS RE-TESTED AND THE REPEAT RESULTS WERE REPRODUCIBLE WITHIN THE SAME CLINICAL RANGE, HOWEVER, COLLECTIVELY THE RESULTS EXCEEDED THE LABELED ACCUTNI ASSAY PRECISION CLAIMS. THE SAMPLE WAS SUBSEQUENTLY REPEAT TESTED IN DUPLICATE ON AN ALTERNATE INSTRUMENT, WHERE REPRODUCIBLE RESULTS WERE GENERATED THAT WERE REGARDED AS VALID. ONE OF THE RESULTS GENERATED FROM THE ALTERNATE INSTRUMENT WAS REPORTED OUTSIDE OF THE LABORATORY. NONE OF THE IMPRECISE ACCUTNI RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THERE WAS NO DEATH, SERIOUS INJURY OR MODIFICATION TO PATIENT TREATMENT ASSOCIATED OR ATTRIBUTED TO THIS EVENT. THE SAMPLE WAS COLLECTED IN A LITHIUM HEPARIN GEL SEPARATOR TUBE AND CENTRIFUGED FOR EIGHT MINUTES AT THREE THOUSAND FIVE HUNDRED REVOLUTIONS PER MINUTE AT ROOM TEMPERATURE PRIOR TO TESTING. THE VOLUME AND THE VISIBLE APPEARANCE OF THE SAMPLE ARE UNKNOWN. THE SAMPLE WAS STORED AT ROOM TEMPERATURE AND WAS PROCESSED THROUGH THE CLOSED TUBE ALIQUOTER AND TESTED FROM THE PRIMARY TUBE. ACCUTNI QUALITY CONTROL (QC) RESULTS ON THE DATE OF THE EVENT PERFORMED WITHIN THE CUSTOMER'S ESTABLISHED LIMITS. BECKMAN COULTER INC. ASSESSMENT OF CUSTOMER SUPPLIED INSTRUMENT/ASSAY PERFORMANCE DATA INDICATED THAT ROUTINE SYSTEM CHECKS PERFORMED PRIOR TO THE EVENT MET INSTRUMENT SPECIFICATIONS. THERE WERE NO ERRORS POSTED TO THE EVENT LOG IN ASSOCIATION WITH THIS EVENT. THE ACCUTNI REAGENT AND CALIBRATOR LOTS ASSOCIATED WITH THIS EVENT WERE 122818 AND 120792 RESPECTIVELY. NO SPECIFIC PATIENT INFORMATION WAS PROVIDED BY THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON LXI 725 SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | MMI | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |