FDA Adverse Event Malfunction Summary report: N

CANCELLOUS SCREW ?4.0X34MM

MDR report key: 2685225 · Received July 31, 2012

Report

Report Number
8031020-2012-00169
Event Type
Malfunction
Date Received
July 31, 2012
Date of Event
July 5, 2012
Report Date
July 5, 2012
Manufacturer
STRYKER TRAUMA AG
Product Code
HWC
PMA / PMN Number
K050512
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS DISCARDED BY THE HOSPITAL. IF ADD¿L INFO IS RECEIVED IT WILL BE SUBMITTED ON A SUPPLEMENTAL REPORT. SAME PT EVENT AS MDR: 8031020-2012-00170.

Description of Event or Problem · 1

IT WAS REPORTED, ¿PT ADMITTED FOR TIBIAL PLATEAU FRACTURE. PROXIMAL LATERAL TIBIA PLATES WERE SHIPPED IN FROM (B)(4) ON DAY OF EVENT WITH CORRECT INSTRUMENTATION AND SCREW CADDY. INSTRUMENTS AND IMPLANTS WERE TAKEN THROUGH STERILE PROCESSING DEPT. AND STERILIZED ACCORDINGLY FOR CASE. INCISION WAS MADE TO EXPOSE THE JOINT AND PROXIMAL LATERAL TIBIA PLATE WAS FIXATED WITH PINS AFTER FRACTURE REDUCTION. FIRST SCREW PLACED INTO METAPHYSIS AND SURGEON CONTINUED BY FIXATING SHAFT SCREWS. NEUTRAL DRILL GUIDE PLACED INTO NON-LOCKING SCREW HOLE. BONE WAS DRILLED BI-CORTICALLY AND SCREW INSERTED FIRST BY HAND, FOLLOWED BY POWER AT SLOW SPEED. SPEED OF DRILL ACCELERATED, BREAKING SCREW APPROX. 3-4 THREADS BELOW HEAD OF SCREW. SCREW WAS LEFT IN SHAFT AS THE SURGEON CONTINUED ON TO NEXT DISTAL SCREW HOLE. NEUTRAL DRILL GUIDE PLACED INTO NON-LOCKING SCREW HOLE AND DRILLED BI-CORTICALLY, THEN OVER DRILLED WITH 3.5 DRILL. INSERTION OF SECOND SCREW STARTED BY HAND AND CONTINUED BY HAND. AS SECOND SCREW SEATED, THE SCREW HEAD BROKE OFF APPROX. 3-4 THREADS BELOW THE HEAD OF THE SCREW, AGAIN LEAVING THE SCREW IN THE BONE. SURGEON THEN MOVED ON TO FIXATE SCREWS IN REMAINING LOCKING AND NON-LOCKING SHAFT HOLES. PROXIMAL LOCKING SCREWS WERE THEN PLACED, HOLDING FRACTURE REDUCED. SURGEON WAS THEN SATISFIED WITH REDUCTION AND ELECTED NOT TO ATTEMPT TO RETRIEVE BROKEN SCREWS. SURGEON THEN ELECTED TO CLOSE.¿

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CANCELLOUS SCREW ?4.0X34MM IMPLANT HWC STRYKER TRAUMA AG NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNK Other