FDA Adverse Event Malfunction Summary report: N

COMPOUNDERS

MDR report key: 2683654 · Received August 6, 2012

Report

Report Number
6000001-2012-13020
Event Type
Malfunction
Date Received
August 6, 2012
Date of Event
May 10, 2012
Report Date
May 10, 2012
Manufacturer
BAXTER HEALTHCARE - BUFFALO GROVE
Product Code
NEP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS IS A CLASS II 510(K) EXEMPT DEVICE WITH THE ADDITIONAL 510(K) OF K961008. THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED. (B)(4). THE CAUSE WAS DETERMINED TO BE A BENT UMBILICAL CABLE. TO CORRECT THE CONDITION, THE UMBILICAL CABLE WAS REPLACED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, AN AUTOMIX COMPOUNDER WAS FOUND TO HAVE A BENT 10 FOOT 25 CONDUCTOR MOLDED UMBILICAL CABLE ASSEMBLY. "THIS WAS NOT REPORTED BY THE CUSTOMER. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATION WITH THIS EVENT." NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COMPOUNDERS SYSTEM/DEVICE, PHARMACY COMPOUNDING NEP BAXTER HEALTHCARE - BUFFALO GROVE