COMPOUNDERS
Report
- Report Number
- 6000001-2012-12981
- Event Type
- Malfunction
- Date Received
- August 3, 2012
- Date of Event
- July 11, 2012
- Report Date
- July 11, 2012
- Manufacturer
- BAXTER HEALTHCARE - BUFFALO GROVE
- Product Code
- NEP
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
(B)(4). THIS IS A CLASS II 510(K) EXEMPT DEVICE WITH THE ADDITIONAL 510(K) OF K955622. THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED. THIS IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE CAUSE WAS DETERMINED TO BE DAMAGE TO THE UMBILICAL CABLE. TO CORRECT THE CONDITION, THE UMBILICAL CABLE WAS REPLACED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.
DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, AN AUTOMIX COMPOUNDER WAS FOUND WITH A DAMAGED UMBILICAL CABLE ASSEMBLY. THIS WAS NOT REPORTED BY THE CUSTOMER. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COMPOUNDERS | SYSTEM/DEVICE, PHARMACY COMPOUNDING | NEP | BAXTER HEALTHCARE - BUFFALO GROVE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |