FDA Adverse Event Malfunction Summary report: N

BIPOL LEAD MODEL 300

MDR report key: 2680340 · Received June 5, 2007

Report

Report Number
1644487-2007-00766
Event Type
Malfunction
Date Received
June 5, 2007
Date of Event
January 1, 2007
Report Date
May 7, 2007
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PT X-RAYS ASSESSED BY MFR. LEAD DISCONTINUITY NOTED DURING REVIEW OF X-RAYS. DEVICE MALFUNCTION IS SUSPECTED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.

Description of Event or Problem · 1

RPTR INDICATED THAT PT PRESENTED FOR ROUTINE OFFICE WHERE HIGH LEAD IMPEDANCE WAS OBTAINED DURING SYSTEM DIAGNOSTIC TESTING INDICATING A POSSIBLE LEAD MALFUNCTION. PT X-RAY WERE SUBSEQUENTLY TAKEN AND SENT IN TO MFR FOR ASSESSMENT. REVIEW OF X-RAYS YIELDED A POSSIBLE LEAD DISCONTINUITY, WHICH WAS NOTED TO BE ADJACENT TO ONE OF THE PT'S TIE-DOWNS. REVISION SURGERY IS LIKELY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BIPOL LEAD MODEL 300 LYJ CYBERONICS, INC. 300-20 5549

Patients

Seq Age Sex Outcome Treatment
1 58 YR