FDA Adverse Event
Malfunction
Summary report: N
BIPOL LEAD MODEL 300
MDR report key: 2680340
·
Received June 5, 2007
Report
- Report Number
- 1644487-2007-00766
- Event Type
- Malfunction
- Date Received
- June 5, 2007
- Date of Event
- January 1, 2007
- Report Date
- May 7, 2007
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PT X-RAYS ASSESSED BY MFR. LEAD DISCONTINUITY NOTED DURING REVIEW OF X-RAYS. DEVICE MALFUNCTION IS SUSPECTED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
Description of Event or Problem · 1
RPTR INDICATED THAT PT PRESENTED FOR ROUTINE OFFICE WHERE HIGH LEAD IMPEDANCE WAS OBTAINED DURING SYSTEM DIAGNOSTIC TESTING INDICATING A POSSIBLE LEAD MALFUNCTION. PT X-RAY WERE SUBSEQUENTLY TAKEN AND SENT IN TO MFR FOR ASSESSMENT. REVIEW OF X-RAYS YIELDED A POSSIBLE LEAD DISCONTINUITY, WHICH WAS NOTED TO BE ADJACENT TO ONE OF THE PT'S TIE-DOWNS. REVISION SURGERY IS LIKELY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BIPOL LEAD MODEL 300 | LYJ | CYBERONICS, INC. | 300-20 | 5549 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |