FDA Adverse Event Malfunction Summary report: N

I-STAT ACT KAOLIN CARTRIDGE

MDR report key: 2678722 · Received July 27, 2012

Report

Report Number
2245578-2012-00371
Event Type
Malfunction
Date Received
July 27, 2012
Date of Event
June 28, 2012
Report Date
June 28, 2012
Manufacturer
ABBOTT POINT OF CARE, INC.
Product Code
JBP
PMA / PMN Number
K023582
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

ON (B)(6) 2012, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING ACT KAOLIN CARTRIDGES THAT YIELDED A HIGH RATE OF MULTIPLE QUALITY CHECK CODES (QCC) 31, 22, AND 25 DURING PATIENT TESTING. THE CUSTOMER WAS USING MULTIPLE ANALYZERS AND 3 DIFFERENT LOTS OF CARTRIDGES. LOT#: S12048B, YIELD RATE: 75% QCC. LOT#: S12122, YIELD RATE: 10% QCC. LOT#: S12105, YIELD RATE: 60% QCC. LOT#: S12122B, MANUFACTURE DATE: 05/2012, EXPIRATION DATE: 10/28/2012. LOT#: S12105, MANUFACTURE DATE: 04/2012, EXPIRATION DATE: 09/28/2012. BASED ON THE INFORMATION AVAILABLE AT THE TIME, THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT. YIELD RATES REPORTED BY THE CUSTOMER ARE UNCONFIRMED. AT THIS TIME ABBOTT POINT OF CARE HAS NO REASON TO BELIEVE THAT A MALFUNCTION EXISTS. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 I-STAT ACT KAOLIN CARTRIDGE ACT KAOLIN CARTRIDGE JBP ABBOTT POINT OF CARE, INC. NA S112048B

Patients

Seq Age Sex Outcome Treatment
1 UNK