FDA Adverse Event Malfunction Summary report: N

POWERED WHEELCHAIR

MDR report key: 2676686 · Received July 31, 2012

Report

Report Number
3004493922-2012-00253
Event Type
Malfunction
Date Received
July 31, 2012
Report Date
July 31, 2012
Manufacturer
INVACARE SUZHOU
Product Code
ITI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

(B)(4) - NO RMA HAS BEEN INITIATED FOR THIS ISSUE. HOWEVER, A QUOTE FOR THE RMA HAS BEEN PROVIDED QT (B)(4). MODEL M51PSEMIBLUE, SERIAL NUMBER/DATE CODE (B)(4) IS APPROXIMATELY 4 MONTHS OLD. THE OWNER'S MANUAL PART NUMBER 1125085 REV J (OCT-08) WAS ISSUED WITH THIS DEVICE. THE OWNER'S MANUAL IS ALSO FOUND ON-LINE AT INVACARE.COM. IT IS UNKNOWN IF THE CONSUMER HAS FULLY READ AND UNDERSTANDS THE OWNER'S MANUAL. DOCUMENTATION PROVIDES WARNINGS, CAUTIONS, AND INSTRUCTIONS FOR SAFELY USING THE DEVICE. IF THE CONSUMER DOES NOT UNDERSTAND THE WRITTEN WARNINGS, CAUTIONS OR INSTRUCTIONS THEN THEY SHOULD CONTACT INVACARE. THE CONSUMER'S AGE, HEIGHT AND WEIGHT ARE UNKNOWN. THE CONSUMER'S MEDICAL CONDITION, STABILITY AND MEDICATION REGIMEN ARE UNKNOWN. THE CONSUMER'S TECHNIQUE WHILE USING THE DEVICE IS UNKNOWN. THE MAINTENANCE HISTORY OF THE DEVICE IS UNKNOWN.

Description of Event or Problem · 1

DEALER STATES THAT "ON SN (B)(4) THAT THE JOYSTICK IS SHORTING OUT AND THE DEALER STATES THE CHAIR IS SLUGGISH. THE DEALER STATES HE SWAPPED OUT THE JOYSTICK FOR ONE OF THEIR STOCK JOYSTICKS AND THERE IS NO ISSUE WHEN HE DOES THAT." QT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 POWERED WHEELCHAIR 890.3860 ITI INVACARE SUZHOU M51PSEMIBLUE

Patients

Seq Age Sex Outcome Treatment
1 79 Other