FDA Adverse Event Injury Summary report: N

FREESTYLE FREEDOM LITE

MDR report key: 2675904 · Received July 31, 2012

Report

Report Number
2954323-2012-06427
Event Type
Injury
Date Received
July 31, 2012
Date of Event
July 16, 2012
Report Date
July 16, 2012
Product Code
NBW
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

REQUESTED PRODUCT WAS NOT RECEIVED FOR INVESTIGATION. RETAINED TEST STRIP SAMPLES FROM THE SAME LOT REPORTED BY THE CUSTOMER (B)(4) WERE TESTED WITH CONTROL SOLUTION INSTEAD. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED. THE COMPLAINT IS NOT CONFIRMED.

Additional Manufacturer Narrative · 1

THIS IS AN INITIAL REPORT. THE PRODUCTS HAVE BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. (B)(4) IT SHOULD BE NOTED: CUSTOMER DECLINED TO PROVIDE HIS WEIGHT. ADDITIONALLY, THE ACTUAL DATE WHEN THE EVENT OCCURRED IS UNKNOWN. THE DATE LISTED IN THIS REPORT IS THE DATE WHEN ABBOTT DIABETES CARE, INC. BECAME AWARE OF THE EVENT.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT APPROXIMATELY 8 WEEKS PRIOR TO CALLING CUSTOMER SERVICE ON (B)(6) 2012 WHILE ON VACATION HE PERFORMED A TEST USING HIS ADC BLOOD GLUCOSE METER AND RECEIVED A READING OF 12 MMOL/L (216 MG/DL), WHICH WAS HIGHER THAN HE FELT. CUSTOMER FURTHER REPORTED THAT 30 MINUTES LATER HE BEGAN TO EXPERIENCE "SHAKING". PARAMEDICS WERE CALLED AND A NEW READING OF 63 MG/DL (3.5 MMOL/L) WAS OBTAINED. IT IS UNKNOWN IF THIS NEW READING WAS OBTAINED ON THE ADC BLOOD GLUCOSE METER OR THE PARAMEDIC'S UNKNOWN BRAND OF METER. CUSTOMER WAS TREATED WITH GLUCOSE VIA INTRAVENOUS INFUSION AND TRANSPORTED TO A LOCAL HEALTHCARE FACILITY. AT THE HOSPITAL CUSTOMER WAS DIAGNOSED WITH HYPOGLYCEMIA AND TREATED WITH ADDITIONAL GLUCOSE VIA IV. CUSTOMER DID NOT TAKE ANY ACTIONS BASED UPON THE INITIAL READING OF 12 MMOL/L (216 MG/DL), AND NOTED JUST TAKING HIS "USUAL (UNSPECIFIED) MEDICATIONS". THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FREESTYLE FREEDOM LITE BLOOD GLUCOSE MONITORING SYSTEM NBW 1258960

Patients

Seq Age Sex Outcome Treatment
1 73 YR Hospitalization| R