ACCU-TEMP CAUTERY
Report
- Report Number
- 1211998-2012-00003
- Event Type
- Malfunction
- Date Received
- July 25, 2012
- Date of Event
- July 10, 2012
- Report Date
- July 23, 2012
- Manufacturer
- BEAVER VISITEC INTERNATIONAL, INC.
- Product Code
- HAM
- PMA / PMN Number
- K833648
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHARMACIST
Narratives
PRODUCT HAS BEEN RETURNED, BUT IT APPEARS TO BE A BVI MANUFACTURED UNIT. THIS IS NOT CONSISTENT WITH THE LOT NUMBER REPORTED. THIS PRODUCT LINE WAS RECENTLY ACQUIRED BY MEDTRONIC AND IS NOW MANUFACTURED AT BEAVER VISITEC INTERNATIONAL. HOWEVER, THE REPORTED LOT NUMBER OF THIS PRODUCT SIGNIFIES THAT IT WAS MANUFACTURED BY MEDTRONIC PRIOR TO THIS ACQUISITION. INVESTIGATION IS IN PROGRESS, WHICH INCLUDES INFORMATION FROM MEDTRONIC REGARDING DATE OF MFR, EXPIRATION DATE, AND COMPLAINT TRENDING. EXPECT A FOLLOW-UP REPORT TO BE PROVIDED ON OR ABOUT (B)(4) 2012.
(B)(4). CUSTOMER REPORTED VERBATIM, "DURING SURGERY, A SWAB CAUGHT FIRE AND BURNED THE SURGEON'S THUMB AND THE SIDE OF THE NOSE AND CHEEK OF THE PATIENT." REPEATED ATTEMPTS TO REACH DR. (B)(6) WERE MADE SINCE (B)(6). HE RETURNED OUR CALL ON (B)(6) 2012. OUR CONVERSATION FOUND THAT SURGERY WAS CONDUCTED TO REMOVE A TUMOR ON THE EYE LID. THE SURGEON DECIDED TO DO A CAUTERIZATION OF A SMALL ARTERY NEXT TO THE LACRIMAL GLAND AND CLOSE TO THE CORNEA. TO PROVIDE EASIER ACCESS HE USED A GAUZE (100% COTTON, NON-WOVEN) PLACING IT UNDER HIS LEFT THUMB TO HELP MAKE THE AREA WHERE THE CAUTERIZATION EASIER TO REACH. AS CAUTERIZATION STARTED, HE FELT A SHARP PAIN IN HIS LEFT THUMB AND STARTED SMELLING SOMETHING BURNING. ALMOST IMMEDIATELY THE PATIENT COMPLAINED INDICATING A SHARP PAIN ON THE SIDE OF HER NOSE. DR. (B)(6) IMMEDIATELY STOPPED AND AS HE REMOVED HIS THUMB AWAY FROM THE PATIENT, HE NOTICED THAT THE GAUZE UNDER HIS THUMB HAD DISINTEGRATED LEAVING A RED MARK ON HIS THUMB AND ON THE PATIENT'S NOSE. THE NEXT DAY THE RED MARK ON DR. (B)(6)'S THUMB WAS GONE. HOWEVER, THE PATIENT WAS STILL SHOWING A RED MARK AND CREAM USED TO TREAT SURFACE BURNS WAS APPLIED. THE PATIENT TODAY DOES NOT SHOW ANY MARK. DR. (B)(6) HAS NOT ISSUES WITH HIS THUMB AND HAS BEEN ABLE TO PERFORM SURGERIES WITHOUT ISSUES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCU-TEMP CAUTERY | ACCU-TEMP CAUTERY 1/2 SHAFT W/FINE TIP | HAM | BEAVER VISITEC INTERNATIONAL, INC. | NA | 71674800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Required Intervention |