FDA Adverse Event Malfunction Summary report: N

MOUNTAINEER OCT MINIPOLYAXIAL SCREWDRIVER

MDR report key: 2672523 · Received July 27, 2012

Report

Report Number
1526439-2012-00159
Event Type
Malfunction
Date Received
July 27, 2012
Date of Event
June 22, 2012
Report Date
September 7, 2012
Manufacturer
DEPUY RAYNHAM
Product Code
LXH
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEPUY SYNTHES SPINE HAS REQUESTED RETURN OF THE DRIVER FOR EVALUATION. A FOLLOW UP REPORT WILL BE FILED UPON RECEIPT OF THE DEVICE AND COMPLETION OF THE EVALUATION.

Additional Manufacturer Narrative · 1

THE COMPLAINT IS NOT CONFIRMED AS THE RETURNED MINIPOLYAXIAL SCREWDRIVER WAS EXAMINED AND FOUND TO FUNCTION AS REQUIRED. THE MINIPOLYAXIAL SCREWDRIVER WAS DESIGNED TO BOTH INSERT AND BACK OUT THE POLYAXIAL SCREWS. A DIFFERENT SCREWDRIVER, THE SUMMIT STYLE MINI-POLYAXIAL SCREWDRIVER, PRODUCT CODE 2883-04-200, IS INTENDED TO BE USED TO HOLD SCREWS STRAIGHT.AT THIS TIME, THE COMPLAINT IS CONSIDERED TO BE CLOSED.

Description of Event or Problem · 1

INTERNATIONAL AFFILIATE REPORTS THAT AFTER INITIAL SPINAL SURGERY WAS PERFORMED ON (B)(6) 2012, THE PATIENT EXPERIENCED NON-DEVICE RELATED PAIN. SURGERY WAS PERFORMED AND PREVIOUSLY IMPLANTED SCREWS WERE REMOVED AND REPLACED. DURING SCREW INSERTION, THE SURGEON NOTICED THAT THE DRIVER AND SCREW DID NOT MAINTAIN ALIGNMENT DURING INSERTION. ANOTHER SCREW WAS TRIED WITH THE DRIVER AND THE CONNECTION BETWEEN THE TIP OF THE DRIVER AND THE SCREW HEAD WAS FOUND TO BE LOOSE. THE SCREW WAS EVENTUALLY PLACED IN BONE WITHOUT INJURY AND SURGERY WAS COMPLETED. HOWEVER, THE PROCEDURE WAS EXTENDED BY THIRTY MINUTES AS A RESULT OF THE DIFFICULTY. THE SURGEON BELIEVES THE DIFFICULTY WAS RELATED TO THE GRIPPING FUNCTION OF THE SCREWDRIVER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MOUNTAINEER OCT MINIPOLYAXIAL SCREWDRIVER ORTHOPEDIC, MANUAL, SURGICAL INSTRUMENT LXH DEPUY RAYNHAM X0609

Patients

Seq Age Sex Outcome Treatment
1 81 YR MOUNTAINEER SCREW, (B)(4)| MOUNTAINEER MINIPOLY DRIVER, (B)(4)