FDA Adverse Event
Other
Summary report: N
CONVERTOR DRAPE (2 DRAPES USED)
MDR report key: 267002
·
Received March 1, 2000
Report
- Report Number
- 267002
- Event Type
- Other
- Date Received
- March 1, 2000
- Date of Event
- February 1, 2000
- Report Date
- February 11, 2000
- Manufacturer
- CONVERTORS
- Product Code
- KDD
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- CA, US
- Reporter Occupation
- INVALID DATA
Narratives
Description of Event or Problem · 1
PT SCHEDULED FOR THROMBECTOMY WITH POSSIBLE REVISION OF GORTEX GRAFT OF LEFT FOREARM. WHILE IN THE OPERATING ROOM, A VALLEYLAB CAUTERY PENCIL WAS USED ON SURGICAL INCISION. THE SURGEON NOTICED SMOKING OF SURGICAL DRAPE AND SUBSEQUENT FLAME OF DRAPE. FLAME IMMEDIATELY EXTINGUISHED. PT SUSTAINED SECOND DEGREE BURN IN THE OPERATING RECORD - THIS RECORDED AS 2CM X 2CM ON LEFT AXILLARY REGION. A TOP DRAPE AND SPLIT DRAPE WERE USED. UNSURE AS TO WHETHER PACK WAS USED FOR CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONVERTOR DRAPE (2 DRAPES USED) | STERILE SURGICAL DRAPE | KDD | CONVERTORS | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Other | VALLEYLAB CAUTERY PENCIL MODEL# E2515HNSB. OXYGEN| THERAPY VIA MASK IN USE AT TIME OF EVENT. |