FDA Adverse Event Malfunction Summary report: N

MEDICHOICE WARMING BLANKET

MDR report key: 2666339 · Received July 18, 2012

Report

Report Number
MW5026247
Event Type
Malfunction
Date Received
July 18, 2012
Date of Event
June 11, 2012
Report Date
July 18, 2012
Manufacturer
*
Product Code
DWJ
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
MI, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

A (B)(6) MALE WAS ADMITTED FOR ELECTIVE AORTIC VALVE REPLACEMENT ON (B)(6) 2012 TO (B)(6) HOSPITAL. THERE WERE NO COMPLICATIONS DURING THE PROCEDURE; HOWEVER, IN RECOVERY STAFF NOTICED THAT THERE WAS A WAFFLE LIKE PATTERN ON THE PT'S BACK THAT WAS THOUGHT TO BE CAUSED BY THE WARMING BLANKET USED DURING SURGERY. BURN LOCATION WAS PT'S UPPER BACK AND BURN PATTERN IS CONSISTENT WITH THE PATTERN WAFFLING ON THE MEDICHOICE HEATING BLANKET. NO BLISTERS OR DRAINAGE NOTED. SURGICAL ADMINISTRATIVE DIRECTOR NOTIFIED AND BIO MED NOTIFIED. IN ADDITION, MANAGER OF MATERIALS MANAGEMENT NOTIFIED. IT WAS LEARNED THAT WE RECENTLY SWITCHED VENDORS AND SURGERY STAFF ARE NOW USING THIS PRODUCT. NOTE IS TAKEN THIS IS THE SECOND INCIDENT OF A SIMILAR NATURE NOTED WITH THE USE OF THIS NEWLY ACQUIRED PRODUCT. USE OF SAID PRODUCT CEASED AFTER THIS EVENT AND SEARCH COMMENCED TO OBTAIN DIFFERENT PRODUCT. EVENT ABATED AFTER USE STOPPED: YES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEDICHOICE WARMING BLANKET WARMING BLANKET DWJ * * *

Patients

Seq Age Sex Outcome Treatment
1 63 YR