FDA Adverse Event
Malfunction
Summary report: N
BAYER IMMUNO I
MDR report key: 266574
·
Received March 2, 2000
Report
- Report Number
- MW1018298
- Event Type
- Malfunction
- Date Received
- March 2, 2000
- Date of Event
- February 29, 2000
- Report Date
- February 29, 2000
- Manufacturer
- BAYER CORP. DIAGNOSTICS DIVISION
- Product Code
- JJC
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
THE LABORATORY HAS 2 BAYER IMMUNO I ANALYZERS WHICH PERFORM IMMUNOASSAYS, FERTILITY TESTS AND CARDIAC MARKERS. DUE TO THE LEAP YEAR DATE AND THE LIS INTERFACE FORMAT UTILIZED, THE COMPUTER SYSTEMS ON THESE ANALYZERS LOCKED UP RENDERING THEM INOPERABLE. TESTING WAS DELAYED APPROXIMATELY 3 HOURS WHILE BAYER ASSISTED FACILITY IN BY-PASSING THE PROBLEM IN ORDER TO PERFORM TESTING ON THESE ANALYZERS. THE LIS DOWNLOAD FUNCTION WAS INACTIVATED, THE INSTRUMENTS WERE REBOOTED AND BECAME OPERABLE. BAYER IS STILL WORKING ON FIXING THIS PROBLEM BUT TESTING CAN BE PERFORMED WITHOUT THE LIS DOWNLOAD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BAYER IMMUNO I | IMMUNOASSAY ANALYZER | JJC | BAYER CORP. DIAGNOSTICS DIVISION | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |