TANGO OPTIMO
Report
- Report Number
- 9610824-2012-00114
- Event Type
- Malfunction
- Date Received
- July 23, 2012
- Date of Event
- June 28, 2012
- Report Date
- July 23, 2012
- Manufacturer
- BIO-RAD MEDICAL DIAGNOSTICS GMBH
- Product Code
- KSZ
- PMA / PMN Number
- BK080013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THIS IS OUR COMBINED INITIAL AND FINAL REPORT ON THIS INCIDENT.
THE CUSTOMER REPORTED THAT A SAMPLE YIELDED IN A SOFTBANK SCREEN A 4+ RESULT IN THE A SPOT AND THE CORD TYPE WAS INTERPRETED AS "POS POS" WHILE TESTING ON THE TANGO YIELDED CLEARLY AN O POS RESULT. THE CUSTOMER CAUGHT THE ANOMALY BEFORE THERE WAS AN ADVERSE RESULT. A REVIEW OF THE TANGO'S RESULT SHEETS DELIVERED CLEARLY SHOW A NEGATIVE REACTION OF THE SAMPLE WITH ANTI-A WHEREAS THE RESULT PRINTOUT FROM SOFTBANK SHOWS A 4+. SINCE THERE IS NO SAMPLE ID VISIBLE ON THE SOFTBANK PRINTOUT (WINDOW ABOVE WINDOW), THERE IS NO EVIDENCE THAT THE DISCREPANT RESULTS ARE ASSIGNED TO THE SAME INDIVIDUAL. BIO-RAD US TECH SUPPORT WAS NOTIFIED ABOUT THE PROPOSAL OF A WORKAROUND SOLUTION GIVEN BY SCC (VENDOR OF SOFTBANK) UNTIL A SUSTAINED RESOLUTION IS AVAILABLE. TAKEN TOGETHER NO FALSE RESULTS HAVE BEEN OBTAINED WITH TANGO OPTIMO S/N (B)(4). PROVIDED THAT DISCREPANT RESULTS HAVE REALLY BEEN ASSIGNED TO AN IDENTICAL PATIENT (NO EVIDENCE) THE ISSUE AROSE DUE TO MIXUP OF THE ORDER OF RESULTS FROM A PARTIALLY REORDERED SAMPLE DURING TRANSFER TO THE HOST SYSTEM VIA SOFTBANK. PRELIMINARY INFORMATION CONCLUDE THAT REPROGRAMMING OF SOFTBANK SHALL PREVENT THE PROBLEM AFTER IMPLEMENTATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TANGO OPTIMO | AUTOMATED BLOOD BANK ANALYSER SYSTEM | KSZ | BIO-RAD MEDICAL DIAGNOSTICS GMBH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |