FDA Adverse Event Malfunction Summary report: N

TANGO OPTIMO

MDR report key: 2664379 · Received July 23, 2012

Report

Report Number
9610824-2012-00114
Event Type
Malfunction
Date Received
July 23, 2012
Date of Event
June 28, 2012
Report Date
July 23, 2012
Manufacturer
BIO-RAD MEDICAL DIAGNOSTICS GMBH
Product Code
KSZ
PMA / PMN Number
BK080013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS IS OUR COMBINED INITIAL AND FINAL REPORT ON THIS INCIDENT.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT A SAMPLE YIELDED IN A SOFTBANK SCREEN A 4+ RESULT IN THE A SPOT AND THE CORD TYPE WAS INTERPRETED AS "POS POS" WHILE TESTING ON THE TANGO YIELDED CLEARLY AN O POS RESULT. THE CUSTOMER CAUGHT THE ANOMALY BEFORE THERE WAS AN ADVERSE RESULT. A REVIEW OF THE TANGO'S RESULT SHEETS DELIVERED CLEARLY SHOW A NEGATIVE REACTION OF THE SAMPLE WITH ANTI-A WHEREAS THE RESULT PRINTOUT FROM SOFTBANK SHOWS A 4+. SINCE THERE IS NO SAMPLE ID VISIBLE ON THE SOFTBANK PRINTOUT (WINDOW ABOVE WINDOW), THERE IS NO EVIDENCE THAT THE DISCREPANT RESULTS ARE ASSIGNED TO THE SAME INDIVIDUAL. BIO-RAD US TECH SUPPORT WAS NOTIFIED ABOUT THE PROPOSAL OF A WORKAROUND SOLUTION GIVEN BY SCC (VENDOR OF SOFTBANK) UNTIL A SUSTAINED RESOLUTION IS AVAILABLE. TAKEN TOGETHER NO FALSE RESULTS HAVE BEEN OBTAINED WITH TANGO OPTIMO S/N (B)(4). PROVIDED THAT DISCREPANT RESULTS HAVE REALLY BEEN ASSIGNED TO AN IDENTICAL PATIENT (NO EVIDENCE) THE ISSUE AROSE DUE TO MIXUP OF THE ORDER OF RESULTS FROM A PARTIALLY REORDERED SAMPLE DURING TRANSFER TO THE HOST SYSTEM VIA SOFTBANK. PRELIMINARY INFORMATION CONCLUDE THAT REPROGRAMMING OF SOFTBANK SHALL PREVENT THE PROBLEM AFTER IMPLEMENTATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TANGO OPTIMO AUTOMATED BLOOD BANK ANALYSER SYSTEM KSZ BIO-RAD MEDICAL DIAGNOSTICS GMBH

Patients

Seq Age Sex Outcome Treatment
1