FDA Adverse Event Injury Summary report: N

VANGUARD BREAST MRI AUXILIARY TABLE W/8CH COIL

MDR report key: 2663058 · Received June 5, 2009

Report

Report Number
3005748597-2009-00001
Event Type
Injury
Date Received
June 5, 2009
Date of Event
June 3, 2009
Report Date
June 5, 2009
Manufacturer
SENTINELLE MEDICAL, INC.
Product Code
MOS
PMA / PMN Number
K060873
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
UT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

DURING ROUTINE BREAST SCREENING MRI EXAM, PATIENT COMPLAINED OF BURNING SENSATION AT STERNUM. AFTER LAST SCAN, PATIENT WAS REMOVED FROM SCANNER, AND EXAMINED BY THE CHARGE TECHNOLOGIST, WHO REPORTED REDNESS "LIKE SUNBURN" CONFINED TO THE PATIENT'S STERNUM. RADIOLOGIST EXAMINED THE PATIENT LATER THE SAME DAY AND SAID REDNESS HAD LARGELY RESOLVED AND HE COULDN'T TELL WHETHER THERE WAS A BURN. MANUFACTURER (SENTIELLE MEDICAL INC) ASKED THE CLINIC TO DISCONTINUE USE OF THE DEVICE PENDING INVESTIGATION, WHICH IS ONGOING. AT THIS TIME THERE IS NO EVIDENCE OF ABNORMAL OPERATION, OR PAST HEATING OF THE DEVICE. IT IS QUITE POSSIBLE THAT THE REDNESS AND BURNING SENSATION WAS CAUSED BY PATIENT WEIGHT BORNE BY STERNUM SUPPORT. SENTINELLE MEDICAL INC. IS CONTINUING TO INVESTIGATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VANGUARD BREAST MRI AUXILIARY TABLE W/8CH COIL VANGUARD FOR GE MRI MOS SENTINELLE MEDICAL, INC. 4000001-12

Patients

Seq Age Sex Outcome Treatment
1 52 YR