TRIAGE CARDIO PROFILER PANEL
Report
- Report Number
- 2027969-2012-01063
- Event Type
- Injury
- Date Received
- July 12, 2012
- Date of Event
- May 29, 2012
- Report Date
- July 12, 2012
- Manufacturer
- ALERE SAN DIEGO, INC
- Product Code
- MMI
- PMA / PMN Number
- 030286
- Removal / Correction Number
- NI
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
PT'S SAMPLE COLLECTED ON (B)(6) 2012, WHICH GAVE A NEGATIVE TNI RESULT WAS RETURNED FOR INVESTIGATION. PRODUCT SUPPORT TESTED RETURNED SAMPLE AND 10 "NORMAL" (APPARENTLY HEALTHY) WHOLE BLOOD EDTA DONORS ON RETAINED LOTS K50690 AND W50654 FOR OBSERVATIONS OF TNI RECOVERY. PRODUCT SUPPORT HAD ADD'L RETENTION INVENTORY OF LOT 250654 AND CONDUCTED FURTHER TESTING ON THIS LOT. CONCLUSION: NO DISCREPANT OR CORRELATION ISSUES WERE OBSERVED WITH WHOLE BLOOD DONORS OR RETURNED PT SAMPLE DURING IN-HOUSE TESTING WITH RETAINED LOTS K50690 AND W50654. TESTING OF BOTH LOTS ON TRIAGE METER GAVE ELEVATED TNI RESULTS, WHICH WERE VERIFIED ON ACCESS. CUSTOMER'S COMPLAINT WAS NOT REPLICATED DURING INVESTIGATION. F/U WITH CUSTOMER REVEALED THAT CUSTOMER BELIEVES THAT THE SECOND TRIAGE TNI RESULT OF <0.05 WAS THE FALSE RESULT. THERE WAS NO CORONARY SPASM DURING ANGIOGRAM AND THIS WOULD HAVE BEEN SEEN. ACCORDING TO CUSTOMER, PT MIGHT HAVE HAD A HEALED PLAQUE RUPTURE WITH A TRANSIENT CORONARY THROMBUS, A MYOCARDITIS. LAB TNI WAS 60 ON THE SAME SAMPLE AS THE NEGATIVE TNI. AS OF (B)(6) 2012, THIS IS THE ONLY COMPLAINT AGAINST LOTS K56090 AND W50654. COMPLAINT ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING. NO CORRECTIVE ACTION REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.
CALLER REPORTED DISCREPANT TROPONIN (TNI) RESULTS ON THE TRIAGE CARDIO PROFILER PANEL. A (B)(6) FEMALE PT PRESENTED AT 10:00 HOURS ON (B)(6) 2012, COMPLAINING OF A MILD PAIN IN THE MIDDLE OF HER BACK. AT 13:15 HOURS, CENTRAL CHEST PAIN WAS RADIATING FROM THE BACK ASSOCIATED WITH NAUSEA, SWEATING, CLAMMINESS AND PINS AND NEEDLES DOWN THE RIGHT ARM. RESULTS AS FOLLOWS: BLOOD DRAW AT 22:05 HOURS ON (B)(6) 2012, LOT W50654: TNI = 0.18; CK-MB = 2.1; MYO = 31.4, TREATED AS ACS. PT SENT TO CATHETER LAB ON (B)(6) 2012 AND CORONARIES WERE NORMAL. EDTA SAMPLE WAS TAKEN ON (B)(6) 2012 AT 15:50 HOURS, LAB TNI = 652, LOT K50690: TNI = <0.05; CK-MB = 1.7; MYO = 30.8. EDTA SAMPLE SENT TO LAB FOR ANALYSIS. RESULT = 500. CARDIOLOGIST TREATED PT AS A NON ST ELEVATED MI (NSTEMI) REGARDLESS OF RESULTS ON TRIAGE METER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRIAGE CARDIO PROFILER PANEL | CARDIAC MARKER PANEL | MMI | ALERE SAN DIEGO, INC | 97100CP | K50690 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR | Other |