FDA Adverse Event Injury Summary report: N

DA VINCI

MDR report key: 2656004 · Received July 11, 2012

Report

Report Number
MW5026151
Event Type
Injury
Date Received
July 11, 2012
Date of Event
June 25, 2012
Report Date
July 3, 2012
Manufacturer
MICROTEK MEDICAL INC
Product Code
NAY
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
TX, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

PT TAKEN TO OPERATING ROOM FOR LAPAROSCOPIC CHOLECYSTECTOMY BY DR. (B)(6). DURING USE OF ELECTROCAUTERY PROBE, THE TIP OF ELECTROCAUTERY PROBE TWISTED OFF FELL OFF OF THE CAUTERY INSTRUMENT AND WAS NOT EASILY IDENTIFIABLE BUT IT WAS THEN IN PT'S ABDOMEN. FLUOROSCOPY WAS BROUGHT IN AND THE TIP OF THE PROBE WAS IDENTIFIED BEHIND THE LIVER. EXTENSIVE EXPLORATION WAS NECESSARY AND AFTER 2.5 HRS THE PROBE WAS FINALLY IDENTIFIED AND IT WAS REMOVED THROUGH THE UMBILICAL PORT ALONG WITH THE GALLBLADDER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DA VINCI CAUTERY HOOK, A-SURE 5MM 10 PACK NAY MICROTEK MEDICAL INC 400156-04 C11283

Patients

Seq Age Sex Outcome Treatment
1 57 YR Required Intervention