TEE/8-3 MHZ TRANSDUCER
Report
- Report Number
- 3032367-2008-00001
- Event Type
- Malfunction
- Date Received
- August 22, 2008
- Date of Event
- July 29, 2008
- Manufacturer
- SONOSITE, INC.
- Product Code
- IYO
- PMA / PMN Number
- K071134
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
ALTHOUGH OUR TEST METHODS HAVE NOT BEEN ABLE TO DUPLICATE THE FAILURE MODE WHEN A TEE TRANSDUCER IS CONNECTED DIRECTLY TO AN M-TURBO SYSTEM WITHOUT THE USE OF A TTC, ENGINEERING ANALYSIS HAS REVEALED A SMALL POTENTIAL FOR THIS SITUATION TO OCCUR. FURTHER TESTING OF THE TEE TRANSDUCER CONNECTED WITH A MICROMAXX SYSTEM VERIFIED THAT THE TEE TRANSDUCERS BEHAVE AS EXPECTED. THE BEHAVIOR ONLY MANIFESTS ITSELF WHEN USED WITH AN M-TURBO SYSTEM, SOFTWARE VERSION 1.1. THE CUSTOMER WAS ADVISED TO NOT USE THEIR TEE TRANSDUCER WITH THEIR M-TURBO SYSTEM. THEY CURRENTLY HAVE MICROMAXX LOANER SYSTEMS. ADDITIONALLY, THEY WERE ADVISED TO TEST THEIR TEE TRANSDUCER TO VERIFY THE ANGLE WAS ADJUSTING AS EXPECTED PRIOR TO USE, AS INSTRUCTED IN THE USER GUIDE.
THE CUSTOMER REPORTED THAT THREE OF THE ANESTHESIOLOGISTS ON SITE HAD REPORTED THAT WHEN ATTEMPTING TO ADJUST THE ANGLE ON THEIR TEE TRANSDUCER, WHEN THE TEE WAS ADJUSTED PAST 94 DEGREES, THE ANGLE WOULD JUMP TO 180 DEGREES INSTEAD OF MOVING IN INCREMENTS OF ONE. WHEN ADJUSTING BACK DOWN FROM 180 DEGREES, THEY FOUND THAT THE ANGLE WOULD AGAIN JUMP DOWN TO 94 DEGREES. ADJUSTMENTS MADE IN THE RANGE BELOW 94 DEGREES MOVED IN SINGLE INCREMENTS AS EXPECTED. THIS ISSUE OCCURRED MID-PROCEDURE ON THREE DIFFERENT PATIENTS. PER THE CUSTOMER, THERE WERE NO INJURIES OR NEGATIVE IMPACT TO THE PATIENT OBSERVED AS A RESULT OF THIS FAILURE. THE CUSTOMER ATTEMPTED TO TROUBLESHOOT THE PROBLEM ON THEIR OWN AND WERE UNABLE TO DUPLICATE THE PROBLEM. THEY CONTACTED SONOSITE TECHNICAL SUPPORT AND FOUND THAT WHEN USED IN PORT 2 OF THEIR TTC, THE TRANSDUCERS OPERATED AS EXPECTED. HOWEVER, THEY WERE ABLE TO DUPLICATE THE ISSUE IN PORTS 1 AND 3.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TEE/8-3 MHZ TRANSDUCER | TRANSESOPHAGEAL ULTRASOUND TRANSDUCER | IYO | SONOSITE, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |