FDA Adverse Event Other Summary report: N

MEAVTRON M2/PRIMUS MID-ENERGY

MDR report key: 2655792 · Received July 5, 2012

Report

Report Number
2910081-2012-05437
Event Type
Other
Date Received
July 5, 2012
Date of Event
June 6, 2012
Report Date
July 2, 2012
Manufacturer
SIEMENS MEDICAL SOLUTIONS USA, INC., RADIATION ONCOLOGY
Product Code
IYE
PMA / PMN Number
K993425
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SIEMENS BECAME AWARE OF THE REPORTED EVENT ON (B)(4) 2012, AND THIS MDR IS BEING MAILED ON 07/03/2012. SIEMENS' INVESTIGATION INTO THE REPORTED ISSUE REVEALS THAT THE MINIMUM DISTANCE OF THE FLAT PANEL DURING GANTRY ROTATION HAS BEEN REDUCED INTENTIONALLY BY A SOFTWARE UPDATE. THE PREVIOUS SOFTWARE VERSIONS HAD LARGER DISTANCES. SAFETY INSTRUCTIONS CONTAINS WARNINGS THAT ADVISE THE USER TO "VERIFY ALL CLEARANCES BEFORE MOVING THE GANTRY WITH THE FLAT PANEL PORTAL IMAGER DEPLOYED" AND TO "VERIFY CLEARANCES BEFORE DEPLOYING OR RETRACTING THE FLAT PANEL PORTAL IMAGER." CUSTOMER ADDRESS: (B)(6).

Description of Event or Problem · 1

SIEMENS WAS NOTIFIED ON (B)(4) 2012 THAT THE FLAT PANEL WILL NO LONGER MOVE TO 160 SID WHILE ROTATING BETWEEN FIELDS. REPORTEDLY, THE CUSTOMER STATED THAT THE FLAT PANEL USED TO AUTOMATICALLY MOVE BACK TO 160 SID AFTER EACH FIELD AND ROTATE. NOW THE FLAT PANEL HAS TO BE SET TO 144.9 SID OR LESS IN ORDER TO MOVE FOR ROTATION BETWEEN FIELDS. THE FLAT PANEL HAS COLLIDED WITH THE TREATMENT TABLE A FEW TIMES DURING PT TREATMENTS. HOWEVER, THERE ARE NO REPORTS OF MISTREATMENT OR INJURY TO A PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEAVTRON M2/PRIMUS MID-ENERGY ACCELERATOR, LINEAR, MEDICAL IYE SIEMENS MEDICAL SOLUTIONS USA, INC., RADIATION ONCOLOGY 1940035 NA

Patients

Seq Age Sex Outcome Treatment
1 SW V4.2| 9.1 MR2