ERBOTOM ICC 200 E
Report
- Report Number
- 9610614-2012-00008
- Event Type
- Other
- Date Received
- July 6, 2012
- Date of Event
- May 31, 2012
- Report Date
- July 6, 2012
- Manufacturer
- ERBE ELEKTROMEDIZIN GMBH
- Product Code
- GEI
- PMA / PMN Number
- K933157
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
THE ESU WAS RETURNED AND THOROUGHLY EVALUATED. THE GENERATOR WAS FOUND TO BE FUNCTIONING AS INTENDED. SPECIFICALLY, A TECHNICAL SAFETY CHECK WAS PERFORMED ON THE UNIT. THIS INCLUDED AN ELECTRICAL SAFETY CHECK, A FUNCTION CHECK OF EACH OF THE EQUIPMENT'S FEATURES, AND A POWER OUTPUT CHECK. ALL FEATURES WERE/ARE FUNCTIONING PROPERLY WITHIN SPECIFICATIONS ON THE ESU. IN ADDITION, NO ANOMALIES WERE FOUND IN THE DEVICE HISTORY RECORD (DHR). IN CONCLUSION, NO EQUIPMENT PROBLEM WAS FOUND THAT WOULD HAVE CAUSED OR ATTRIBUTED TO THE EVENT. THE INVOLVED MEDICAL PERSONNEL ARE BEING MADE AWARE OF THE FINDINGS. TO FURTHER ADDRESS THE ISSUE, ADDITIONAL IN-SERVICE WORK WAS PERFORMED WITH THE INVOLVED MEDICAL PERSONNEL ON 06/29/2012. ADDITIONALLY, THE ACCOUNT WILL BE ADVISED TO CHOOSE AND ADJUST SETTINGS AS NECESSARY TO OBTAIN THE DESIRED TISSUE EFFECT. NO TRENDS HAVE BEEN IDENTIFIED WITH THIS INCIDENT. ERBE USA, INC IS NOW CLOSING THE FILE ON THIS EVENT.
THE ACCOUNT REPORTED THAT THE ELECTROSURGICAL UNIT (ESU/GENERATOR) WAS INVOLVED IN A PT INCIDENT. THE ESU WAS BEING USED TO REMOVE A POLYP. THE SETTINGS WERE CUT 200 WATTS/SOFT COAG 60 WATTS, EFFECT 3. THE SNARE ADHERED TO THE POLYP TISSUE AND THEY WERE UNABLE TO REMOVE IT. THEREFORE, THE SURGERY WAS PERFORMED TO REMOVE THE SNARE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ERBOTOM ICC 200 E | ELECTROSURGICAL GENERATOR | GEI | ERBE ELEKTROMEDIZIN GMBH | ICC 200 E | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR | Required Intervention |