FDA Adverse Event Injury Summary report: N

GORE TAG THORACIC ENDOPROSTHESIS

MDR report key: 2655726 · Received July 3, 2012

Report

Report Number
2017233-2012-00444
Event Type
Injury
Date Received
July 3, 2012
Date of Event
August 1, 2007
Report Date
June 14, 2012
Manufacturer
W.L. GORE & ASSOCIATES
Product Code
MIH
PMA / PMN Number
P040043
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CONCLUSION - THE ROOT CAUSE OF THE GRAFT ERODING THE PREVIOUSLY PLACED DACRON GRAFT IS UNK. FEHRENBACHER, JOHN W AND ROBERT A MCCREADY. "EROSION OF ELEPHANT TRUNK DACRON GRAFT LIMB BY THORACIC ENDOGRAFT CAUSING ACUTE ANEURYSM EXPANSION." JOURNAL OF VASCULAR SURGERY 2009, 49: 491-3.

Description of Event or Problem · 1

AN ARTICLE IN THE JOURNAL OF VASCULAR SURGERY REPORTS A PT WHO WAS IMPLANTED WITH TWO GORE TAG THORACIC ENDOPROSTHESES TO REPAIR A DESCENDING THORACIC AORTIC ANEURYSM AS THE SECOND STAGE OF A HYBRID PROCEDURE. THE PT HAD PREVIOUSLY UNDERGONE REPAIR OF ASCENDING AND AORTIC ARCH ANEURYSMS, WITH AN ELEPHANT TRUNK GRAFT LIMB PLACED IN THE DESCENDING THORACIC AORTA. EIGHT MONTHS AFTER PLACEMENT OF THE THORACIC ENDOGRAFT, THE PT PRESENTED WITH AN ACUTELY EXPANDING AND SYMPTOMATIC THORACIC ANEURYSM. THE PT WAS OPERATED ON URGENTLY. THE PROXIMAL PORTION OF THE ENDOGRAFT HAD ERODED INTO THE PREVIOUSLY PLACED DACRON ELEPHANT TRUNK LIMB. THE PROXIMAL PORTION OF THE ENDOGRAFT WAS REMOVED AND WAS REPLACED WITH A DACRON GRAFT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GORE TAG THORACIC ENDOPROSTHESIS MIH/SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH W.L. GORE & ASSOCIATES

Patients

Seq Age Sex Outcome Treatment
1 45 YR Hospitalization| R