FDA Adverse Event Malfunction Summary report: N

FLEXIVA (TM) HIGH POWER SINGLE-USE LASER FIBER

MDR report key: 2653012 · Received July 12, 2012

Report

Report Number
3005099803-2012-02598
Event Type
Malfunction
Date Received
July 12, 2012
Date of Event
June 19, 2012
Report Date
June 21, 2012
Manufacturer
BOSTON SCIENTIFIC - MARLBOROUGH
Product Code
GEX
PMA / PMN Number
K100078
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

VISUAL ANALYSIS OF THE RETURNED FIBER REVEALED APPROXIMATELY 5.0 MM OF EXPOSED GLASS TIP REMAINING. THE AIMING BEAM WAS NOT EXITING FROM THE DISTAL END OR FROM ANY OTHER LOCATION ON THE BODY OF THE FIBER INDICATING THAT THE FIBER IS BROKEN WITHIN THE CONNECTOR.

Additional Manufacturer Narrative · 1

THE FOLLOWING INFORMATION WAS PROVIDED BY THE CUSTOMER ON (B)(6) 2012: PATIENT IDENTIFIER (B)(6). DATE OF BIRTH (B)(6) 1939. PATIENT SEX FEMALE. DEVICE LOT # - ML00000639, DEVICE EXPIRATION DATE 05/31/2015. ADD'L DEVICE: GRID LASER TYPE (HOLMIUM LASER), THERAPY DATE ((B)(6)2012), (B)(6).

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT DURING A LITHOTRIPSY OF STONE PROCEDURE USING A FLEXIVA 200 LASER FIBER, THE CONNECTOR WENT ON; ALTHOUGH IT APPEARED NOT TO GO ON ALL THE WAY. THE FIBER HEATED UP AND THE ENERGY WAS NO LONGER BEING EMITTED FROM THE OTHER SIDE OF THE FIBER. THE PROCEDURE WAS COMPLETED WITH ANOTHER FLEXIVA 200 LASER FIBER WITHOUT ANY COMPLICATIONS TO THE PATIENT, WHO IS REPORTEDLY "FINE" POST PROCEDURE. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT DURING A LITHOTRIPSY OF STONE PROCEDURE USING A FLEXIVA 200 LASER FIBER, THE CONNECTOR WENT ON; ALTHOUGH IT APPEARED NOT TO GO ON ALL THE WAY. THE FIBER HEATED UP AND THE ENERGY WAS NO LONGER BEING EMITTED FROM THE OTHER SIDE OF THE FIBER. THE PROCEDURE WAS COMPLETED WITH ANOTHER FLEXIVA 200 LASER FIBER WITHOUT ANY COMPLICATIONS TO THE PATIENT, WHO IS REPORTEDLY 'FINE' POST PROCEDURE. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT DURING A LITHOTRIPSY OF STONE PROCEDURE USING A FLEXIVA 200 LASER FIBER, THE CONNECTOR WENT ON; ALTHOUGH IT APPEARED NOT TO GO ON ALL THE WAY. THE FIBER HEATED UP AND THE ENERGY WAS NO LONGER BEING EMITTED FROM THE OTHER SIDE OF THE FIBER. THE PROCEDURE WAS COMPLETED WITH ANOTHER FLEXIVA 200 LASER FIBER WITHOUT ANY COMPLICATIONS TO THE PATIENT, WHO IS REPORTEDLY 'FINE' POST PROCEDURE. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXIVA (TM) HIGH POWER SINGLE-USE LASER FIBER POWERED LASER SURGICAL INSTRUMENT GEX BOSTON SCIENTIFIC - MARLBOROUGH M0068403911 ML00000639

Patients

Seq Age Sex Outcome Treatment
1 73 YR HOLMIUM LASER