FDA Adverse Event Malfunction Summary report: N

PRIM SYM SET 2 CL W/FILTERTEK VALVE

MDR report key: 2652536 · Received July 5, 2012

Report

Report Number
9615050-2012-00740
Event Type
Malfunction
Date Received
July 5, 2012
Date of Event
June 1, 2012
Report Date
June 6, 2012
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
UNK
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER CONTACT INDICATED THAT THE DEVICE WAS DISCARDED; HOWEVER, A REP DEVICE OF THE SAME LIST NUMBER FROM HOSPIRA STOCK WAS EVALUATED. THE DEVICE PASSED TESTING. NO BACKFLOW OF FLUID WAS NOTED. BASED ON THE DATA VERIFIED, HOSPIRA COULD NOT ATTRIBUTE THE ISSUE TO THE DEVICE. THE LIST NUMBER HAS A COMMON DEVICE NAME OF 80FPA AND A 510K OF K103344. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED BACKFLOW OF FLUID FROM THE SECONDARY SOLUTION CONTAINER INTO THE PRIMARY TUBING CONTAINER. THE PRIMARY TUBING SET WAS BEING USED TO DELIVER AN UNSPECIFIED MEDICATION, AT AN UNSPECIFIED RATE, VIA A SYMBIQ PUMP. AT AN UNSPECIFIED TIME, THE MALE ADAPTER OF AN UNSPECIFIED SECONDARY TUBING SET WAS CONNECTED TO THE PROXIMAL CLAVE Y-SITE ON THE PRIMARY TUBING SET FOR PIGGYBACK DELIVERY OF AN UNSPECIFIED CHEMOTHERAPEUTIC AGENT. AFTER AN UNSPECIFIED LENGTH OF TIME, THE NURSE NOTED BACKFLOW OF SOLUTION FROM THE SECONDARY SOLUTION CONTAINER INTO THE PRIMARY TUBING CONTAINER. THERE WERE NO REPORTED ADVERSE PT EFFECTS AND NO REPORTED DELAY OF THERAPY CRITICAL TO THIS PT. NO MEDICAL INTERVENTIONS WERE REPORTED. THOUGH REQUESTED, NO ADDITIONAL INFO WAS PROVIDED INCLUDING IF THE PRIMARY SOLUTION CONTAINER WAS HUNG LOWER THAN THE SECONDARY SOLUTION CONTAINER AND IF THE TUBING SET WAS REPLACED AND THE THERAPY WAS RESUMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PRIM SYM SET 2 CL W/FILTERTEK VALVE UNK FRN HOSPIRA COSTA RICA LTD. NA UNK5H

Patients

Seq Age Sex Outcome Treatment
1 UNK LOT# UNK| UNSPECIFIED SECONDARY TUBING SET: LIST# UNK