LCD WITH TOUCHSCREEN
Report
- Report Number
- 1718850-2012-00716
- Event Type
- Other
- Date Received
- June 27, 2012
- Date of Event
- August 1, 2011
- Report Date
- June 14, 2012
- Manufacturer
- SORIN GROUP DEUTSCHLAND
- Product Code
- DTQ
- PMA / PMN Number
- K071318
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- UNKNOWN
Narratives
SORIN GROUP (B)(4) MANUFACTURERS THE LCD WITH TOUCHSCREEN. THE INCIDENT OCCURRED IN (B)(6). THIS MDR REPORT IS BEING SUBMITTED ON BEHALF OF THE DEVICE MANUFACTURER - SORIN GROUP (B)(4). TO RESOLVE AN FDA INSPECTIONAL OBSERVATION, THE SORIN GROUP (B)(4) MDR PROCEDURE WAS REVISED. AS COMMITTED TO FDA'S OFFICE OF COMPLIANCE, A RETROSPECTIVE REVIEW OF CUSTOMER COMPLAINTS IS BEING PERFORMED AND MDRS WILL BE SUBMITTED IN ACCORDANCE WITH THE REVISED PROCEDURE. THIS MDR IS BEING SUBMITTED BEYOND THE 30 DAY REPORTING REQUIREMENT AS IT WAS IDENTIFIED DURING THE RETROSPECTIVE REVIEW. THE TOUCHSCREEN WAS RETURNED TO SORIN (B)(4) FOR EVALUATION. TESTING FOUND THAT LIQUID HAD INFILTRATED THE TOUCHSCREEN, CAUSING IT TO FAIL. HOW THE LIQUID INFILTRATED THE TOUCHSCREEN IS UNK. NO FURTHER ACTION IS DEEMED NECESSARY.
SORIN GROUP (B)(4) RECEIVED A REPORT THAT THE TOUCHSCREEN WAS UNRESPONSIVE. THERE WAS NO PT INVOLVEMENT.